Radiometer ABL90 FLEX Plus Analyzers Recalled for Measurement Errors
Radiometer Medical ApS is recalling ABL90 FLEX Plus analyzers and hemolyzer units because defective cuvettes cause inaccurate blood measurements.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where a defect causes measurement errors, representing a risk of harm to patients through inaccurate clinical data, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Radiometer Medical ApS is recalling 544 units of ABL90 FLEX Plus analyzers (model no. 393-02) and Spare Part Hemolyzer Units (902-980) with serial numbers R0526 through R0941. The recall covers products distributed worldwide, including the United States.
The recall is due to a defect in the hemolyzer cuvettes, a subcomponent of the optical system used to measure carboxyhemoglobin (ctHb) and its derivatives. A light shunt in the cuvettes prevents some light from passing through the sample, resulting in a perturbed and non-linear absorbance spectrum. This leads to measurement errors in the analyzer's output.
The affected devices are intended for use by trained technologists, nurses, physicians, and therapists in laboratory, near-patient, or point-of-care settings. Users should contact the manufacturer for instructions on how to address the defect.
The recalled product
- Product
- ABL90 FLEX PLUS, model no. 393-02; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
- Manufacturer
- Radiometer Medical ApS
- Hazard
- Affected units
- 544
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- ABL90 FLEX PLUS Model no. 393-092: R0014 - R0195
- and Spare Part Hemolyzer Units (902-980): R0526 - R0941
- Model Number 393-092.
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 2 recalled by Radiometer Medical ApS under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Radiometer Medical ApS
See all →- ModerateRadiometer ABL90 Series Blood Analyzers May Unexpectedly Freeze
FDA (Devices) · 2025-04-30
- HighABL800 FLEX Blood Gas Analyzers May Report Inaccurate pH Results
FDA (Devices) · 2024-07-31
- HighRadiometer AQURE Software Recall Due to Patient Mix-up Risk
FDA (Devices) · 2023-04-26
- HighABL800 Flex Analyzer Recalled for Potential Measurement Bias
FDA (Devices) · 2022-07-13
- HighRadiometer Arterial Blood Sampler recalled due to sterility guarantee loss
FDA (Devices) · 2022-05-25
Same hazard · measurement error
See all →- ModeratePhilips Ingenia Ambition X MRI Systems Stiffness Value Error Recall
FDA (Devices) · 2026-05-06
- ModeratePhilips MR 7700 MRE Upgrade MRI Equipment Stiffness Value Error Risk
FDA (Devices) · 2026-05-06
- HighPhilips Evolution Upgrade 1.5T MRI Stiffness Calculation Error
FDA (Devices) · 2026-05-06
- ModeratePhilips SmartPath to Ingenia Elition X MR Elastography Stiffness Value Errors
FDA (Devices) · 2026-05-06
- ModeratePhilips MR 7700 MR Elastography device stiffness value error recall
FDA (Devices) · 2026-05-06