The Recall Desk

FDA (Devices) recall action 85390

Radiometer Medical ApS recalled 2 products on 2020-06-03

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2020-06-03
Critical
0
Units
1,777

Why these products were recalled

The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2020-06-03

    Radiometer ABL90 FLEX Analyzers Recalled for Hemolyzer Measurement Errors

    Radiometer Medical ApS is recalling ABL90 FLEX analyzers and spare hemolyzer units because defective cuvettes cause light shunts, leading to inaccurate measurement results.

    Product
    ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-06-03

    Radiometer ABL90 FLEX Plus Analyzers Recalled for Measurement Errors

    Radiometer Medical ApS is recalling ABL90 FLEX Plus analyzers and hemolyzer units because defective cuvettes cause inaccurate blood measurements.

    Product
    ABL90 FLEX PLUS, model no. 393-02; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care…