Siemens Dimension Vista Systems Recalled for Potential Sample Dispensing Error
Siemens is recalling Dimension Vista 500 and 1500 systems due to a rare software issue that may cause samples to be dispensed into previously used wells.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source explicitly states the defect occurs under 'specific rare conditions' and reports no illnesses or injuries, fitting the criteria for a moderate severity voluntary precautionary recall.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is conducting a field correction for Dimension Vista 500 and Dimension Vista 1500 systems equipped with software versions 3.5.1 and 3.6. The recall involves 1,937 units distributed worldwide, including the United States and numerous international markets.
The recall is issued because, under specific rare conditions requiring a system reset, there is a potential for a sample to be dispensed into an aliquot well that has previously been used for a quality control (QC) or calibrator product. This error could occur if two samples are dispensed into the same aliquot well, potentially compromising the integrity of the test results.
Healthcare facilities and laboratories using these systems should contact Siemens Healthcare Diagnostics for instructions on the field correction. The source text does not report any illnesses or injuries associated with this defect.
The recalled product
- Product
- Dimension Vista 500 and Dimension Vista 1500 System Aliquot Well Double Dispense with software versions 3.5.1 and 3.6. In vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to…
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 1,937
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Dimension Vista 500 Siemens material numbers: 10284473 (domestic)
- and 10488224 (international)
Distribution
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · sample contamination
See all →- ModerateBD FACS Sample Prep Assistant III Recalled for Probe Wear Issues
FDA (Devices) · 2020-06-17
- ModerateBiomedical Polymers Recalls Hemo-Drop Blood Dispensers Due to Metal Cannula Defect
FDA (Devices) · 2019-11-13
- ModerateBiomedical Polymers Recalls Manual Diff Droppers Due to Cannula Failure
FDA (Devices) · 2019-11-13
- HighSiemens Dimension Vista 1500 Recalled for Potential Sample Contamination
FDA (Devices) · 2018-12-26
- HighSiemens Dimension Vista 500 Recalled for Sample Contamination Risk
FDA (Devices) · 2018-12-26
Similar recalls
- HighSiemens Dimension Vista 500 Recalled for Sample Contamination Risk
FDA (Devices) · 2018-12-26
- HighSiemens Dimension Vista 1500 Recalled for Potential Sample Contamination
FDA (Devices) · 2018-12-26
- HighSiemens Dimension Vista 1500 Lab System Software Defect Recall
FDA (Devices) · 2016-06-08
- HighSiemens Dimension Vista 500 Lab System Software Defect Recall
FDA (Devices) · 2016-06-08
- ModerateSiemens Dimension Vista Diagnostic Vials Recalled for Volume Error
FDA (Devices) · 2013-12-11
- ModerateSiemens Dimension Vista Systems Recalled for Software Processing Errors
FDA (Devices) · 2015-07-15