Siemens Dimension Vista Systems Recalled for Software Processing Errors
Siemens is recalling Dimension Vista Systems due to software issues that may cause samples to stop processing or deliver incorrect reagents, leading to unexpected test results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves software errors leading to incorrect test results, which is a significant quality issue but does not involve direct physical injury or death, fitting the Moderate severity criteria for low-risk contamination or labeling/processing errors.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling 2,315 Dimension Vista Systems used with the Dimension Vista Intelligent Lab System software version 3.6.1. These in vitro diagnostic analyzers were distributed worldwide, including in the United States, Puerto Rico, and numerous international countries.
The recall is due to two software issues. First, samples may stop processing without notification. Second, a timing issue in the Vista 1500 model may cause the wrong reagent to be delivered or no reagent to be delivered at all. These issues may lead to unflagged, unexpectedly low results or results flagged with assay errors.
The affected units are identified by Siemens Material Numbers (SMN) 10284473, 10488224, 10444801, and 10444802. Users of these systems should contact Siemens Healthcare Diagnostics for further instructions on resolving the software issues.
The recalled product
- Product
- Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software version 3.6.1. In vitro diagnostic analyzer.
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 2,315
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- Siemens Material Number (SMN): 10284473
- 10488224
- 10444801
- 10444802
Distribution
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · software error
See all →- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter AU480 Chemistry Analyzer Software Error Recall
FDA (Devices) · 2026-10-07
- HighBard EnCor EnCompass Breast Biopsy System Software Recall
FDA (Devices) · 2026-10-07
- HighPhilips Spectral CT 7500 Software Recall for Radiation Exposure Risk
FDA (Devices) · 2026-10-07
- HighPhilips EPIQ Elite Ultrasound System Recalled for Software Error
FDA (Devices) · 2026-09-30
Similar recalls
- HighSiemens Dimension Vista 1500 Lab System Software Defect Recall
FDA (Devices) · 2016-06-08
- HighSiemens Dimension Vista 500 Lab System Software Defect Recall
FDA (Devices) · 2016-06-08
- ModerateSiemens Dimension Vista Lab Systems Recalled for Reagent Performance Issues
FDA (Devices) · 2014-12-31
- ModerateSiemens Dimension Vista Diagnostic Vials Recalled for Volume Error
FDA (Devices) · 2013-12-11
- ModerateSiemens Dimension Vista Lab Systems Recalled for Potential ECREA Test Inaccuracy
FDA (Devices) · 2017-04-05
- ModerateSiemens Dimension Vista Systems Recalled for Potential Sample Dispensing Error
FDA (Devices) · 2013-09-04