Siemens Dimension Vista 500 Recalled for Sample Contamination Risk
Siemens is recalling Dimension Vista 500 instruments due to a software issue that may cause sample contamination if tests are manually cancelled during aliquoting.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (medical device) where injury has not yet been reported, fitting the 'High' severity criteria for potential harm without confirmed adverse events.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling 1,649 Dimension Vista 500 in vitro diagnostic devices, model no. 10284473. These devices are used to determine analytes in human body fluids by duplicating manual analytical procedures. The recall involves units running Dimension Vista software versions 3.7 and 3.8.1, which were distributed nationwide and worldwide.
The recall is due to a software issue that may result in aliquot well contamination. If an operator manually cancels a test for calibration, quality control, or a patient sample while the instrument is aliquoting, the device may continue to deliver the sample or vial contents to the aliquot well. This can lead to future samples being assigned to the same well and becoming contaminated, or overflow that contaminates surrounding wells.
This issue only occurs when an operator manually cancels a test while the instrument is in the process of aliquoting the sample. Users of these devices should contact Siemens Healthcare Diagnostics for further instructions on how to address the software issue.
The recalled product
- Product
- Dimension Vista 500, model no. 10284473 an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Hazard
- Affected units
- 1,649
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Dimension Vista software versions 3.7 and 3.8.1
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 2 recalled by Siemens Healthcare Diagnostics, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · sample contamination
See all →- ModerateBD FACS Sample Prep Assistant III Recalled for Probe Wear Issues
FDA (Devices) · 2020-06-17
- ModerateBiomedical Polymers Recalls Hemo-Drop Blood Dispensers Due to Metal Cannula Defect
FDA (Devices) · 2019-11-13
- ModerateBiomedical Polymers Recalls Manual Diff Droppers Due to Cannula Failure
FDA (Devices) · 2019-11-13
- HighSiemens Dimension Vista 1500 Recalled for Potential Sample Contamination
FDA (Devices) · 2018-12-26
- HighRoche Recalls Cobas 8100 Reformatter Modules Due to Rack Crash Risk
FDA (Devices) · 2017-04-19
Similar recalls
- ModerateSiemens Dimension Vista Systems Recalled for Potential Sample Dispensing Error
FDA (Devices) · 2013-09-04
- HighSiemens Dimension Vista 500 Lab System Software Defect Recall
FDA (Devices) · 2016-06-08
- HighSiemens Dimension Vista 1500 Lab System Software Defect Recall
FDA (Devices) · 2016-06-08
- ModerateSiemens Dimension Vista Lab Systems Recalled for Reagent Performance Issues
FDA (Devices) · 2014-12-31
- ModerateSiemens Dimension Vista Diagnostic Vials Recalled for Volume Error
FDA (Devices) · 2013-12-11
- HighSiemens Dimension Vista 500 software issue affects test accuracy
FDA (Devices) · 2021-12-08