The Recall Desk
HighFDA (Devices)·Z-2057-2015·Announced 2015-08-05

Hamilton-G5 Ventilators Recalled for Display Freeze Issue

Hamilton Medical is recalling 248 Hamilton-G5 ventilators because the display can freeze, preventing users from operating the device while ventilation continues.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a high-risk medical device where a display freeze prevents user operation, creating a risk of harm, but the source text does not report any specific injuries or illnesses.

Plain-English summary

Hamilton Medical, Inc. is recalling 248 Hamilton-G5 Ventilators with software versions V2.40 and 2.41. The affected devices are used for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The ventilators were distributed nationwide.

The recall is due to customer reports that the ventilator display can freeze. When this occurs, ventilation continues, but the information is no longer displayed on the screen, and the user can no longer operate the device.

Healthcare facilities should contact Hamilton Medical for instructions on how to resolve the issue. The agency has classified this as a Class II recall.

The recalled product

Product
Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.
Affected units
248

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Software version 2.20 and 2.41

Distribution

Distributed nationwide across the United States.