Terumo Sarns Centrifugal Pump Recalled for Pumphead Leaks
Terumo is recalling specific lots of Sarns Disposable Centrifugal Pumps because leaks were detected on the pumphead during set-up and priming.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a high-risk medical device used in cardiopulmonary bypass where a leak could cause harm, but the source text does not report any injuries or illnesses, fitting the 'High' criteria for risk-of-harm products without reported injury.
Plain-English summary
Terumo Cardiovascular Systems Corporation is recalling the Sarns Disposable Centrifugal Pump with X-Coating, model 164275X. This sterile, single-use device is intended for use in cardiopulmonary bypass procedures.
The recall was initiated because leaks were detected on the centrifugal pumphead during the set-up and priming of the bypass circuit. The affected lots are RC07, QM03, QN07, and RA09. The products were distributed nationwide.
Healthcare facilities and users should stop using the affected devices and contact Terumo Cardiovascular Systems Corporation for further instructions or a return.
The recalled product
- Product
- Sarns Disposable Centrifugal Pump with X-Coating, 164275X, is a sterile, single use device. For use in cardiopulmonary bypass procedures.
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- Lot No. RC07
- QM03
- QN07
- and RA09.
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by Terumo Cardiovascular Systems Corporation under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Terumo Cardiovascular Systems Corporation
See all →- HighCDI OneView System BPM Probe temperature sensor measurement inaccuracy
FDA (Devices) · 2025-07-02
- HighTerumo CDI OneView Monitor Probe May Fail to Display Oxygen Saturation Data
FDA (Devices) · 2025-04-09
- HighCardiovascular Procedure Kits Recalled for Pressure Relief Valve Defect
FDA (Devices) · 2024-04-10
- HighCAPIOX FX25 Oxygenator Recalled for Potential Leaks from Improper Bonding
FDA (Devices) · 2023-04-12
- HighTerumo Recalls CAPIOX NX19 Oxygenator Due to Improper Bonding
FDA (Devices) · 2023-04-12
Same hazard · leak
See all →- HighAVID Medical Recalls Halyard LAVH PACK Due to Sterilization Error
FDA (Devices) · 2026-08-26
- HighMedline Convenience Kits with Recalled Swan-Ganz Catheters
FDA (Devices) · 2026-07-08
- HighBicarby dialysate luer-lock connections may leak fluid onto floors
FDA (Devices) · 2026-06-17
- HighSwan-Ganz Pacing Catheter Model D200F7 Recall Due to Leak Risk
FDA (Devices) · 2026-05-27
- HighChemotherapy bag spikes recalled because port weld may separate and leak
FDA (Devices) · 2025-11-05