The Recall Desk

Manufacturer

Terumo Cardiovascular Systems Corporation

204 recalls in our database name Terumo Cardiovascular Systems Corporation as the manufacturing or recalling firm.

What the numbers show

Total
204
Critical
0
Severe
20
Most recent
2025-07-02

Of the 204 recalls from Terumo Cardiovascular Systems Corporation we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 20 (10%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–25 of 204

  • HighFDA (Devices)·Z-2018-2025·2025-07-02

    CDI OneView System BPM Probe temperature sensor measurement inaccuracy

    Terumo Cardiovascular Systems is recalling CDI OneView System BPM Probe units due to thermistor assembly errors that cause inaccurate temperature and blood gas measurements, potentially leading to inappropriate clinical decisions.

    Product
    CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1492-2025·2025-04-09

    Terumo CDI OneView Monitor Probe May Fail to Display Oxygen Saturation Data

    Terumo's CDI OneView Hematocrit/Oxygen Saturation Probe may intermittently fail to display oxygen saturation values during clinical use. When this occurs, the monitor displays dashes and triggers a yellow low-limit alarm.

    Product
    Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
    Category
    Medical Device
    Distribution
    6 states
  • HighFDA (Devices)·Z-1401-2024·2024-04-10

    Cardiovascular Procedure Kits Recalled for Pressure Relief Valve Defect

    Terumo is recalling 20,533 cardiovascular procedure kits due to a pressure relief valve that opens at incorrect pressure (200-300 mmHg instead of 400 mmHg). The defect could cause myocardial tissue damage during cardiac surgery.

    Product
    Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 66312 66342 66388 66650 66852 70472 74307 74365 74366 74430 74475 74477 74502 74773 74873 75065 75157 75195 75409 75410 75473 75533 75534…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1345-2023·2023-04-12

    CAPIOX RX15 Oxygenator units recalled due to improper thermistor bonding

    Terumo is recalling CAPIOX RX15 Hollow Fiber Oxygenators due to improper bonding of the arterial thermistor, which may cause leaks at the connection port.

    Product
    CAPIOX RX15 Hollow Fiber Oxygenator, Catalog Number 3CX*RX15RE30
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1346-2023·2023-04-12

    Terumo Recalls CAPIOX NX19 Oxygenator Due to Improper Bonding

    Terumo Cardiovascular is recalling CAPIOX NX19 oxygenators with improperly bonded arterial thermistors that may leak. Eighty-four units were distributed across the US and internationally.

    Product
    CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1347-2023·2023-04-12

    CAPIOX FX25 Oxygenator Recalled for Potential Leaks from Improper Bonding

    Terumo Cardiovascular Systems is recalling 4,940 CAPIOX FX25 oxygenators because an arterial thermistor component may be improperly bonded to the oxygenator port, potentially causing leaks.

    Product
    CAPIOX FX25 Hollow Fiber Oxygenator, Catalog Numbers 3CX*FX25REC, 3CX*FX25RWC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2350-2021·2021-09-01

    Terumo Advanced Perfusion System Electronic Patient Gas System Recalled for Control Failure

    The Terumo Advanced Perfusion System 1 Electronic Patient Gas System is recalled for intermittent initialization failures that disable the primary control interface, requiring use of backup local controls.

    Product
    Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1789-2021·2021-06-09

    Terumo HX2 and TCM Temperature Management Systems Recalled for Cleaning Protocol Issues

    Terumo recalled the HX2 and TCM temperature management systems used in cardiac surgery due to inability to validate an updated cleaning protocol. Users should discontinue use and properly dispose of affected devices.

    Product
    The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1791-2021·2021-06-09

    Terumo Sarns TCM II cardiac temperature controller cleaning protocol failure

    The Sarns TCM II temperature control system used in cardiac surgery has been recalled due to inability to validate its cleaning protocol. Users are instructed to discontinue use and dispose of all affected devices.

    Product
    The Sarns TCM II (system) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. It also freezes water for an ice supply, monitors temperatures in the patient and…
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-1790-2021·2021-06-09

    Terumo Sarns Temperature Control Devices Recalled for Cleaning Protocol Validation Issues

    Terumo is recalling Sarns Temperature Control Monitor units (1,176 devices) because it cannot validate a cleaning protocol meeting current FDA requirements. Users should discontinue use and dispose of affected devices immediately.

    Product
    The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. The TCM with options will also supply water for cardioplegia…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1367-2021·2021-04-14

    Terumo Sternal Retractor Locking Mechanism Recall for Fracture Risk

    Terumo is recalling 12 sternal retractor devices because the stainless steel locking plates may fracture. The devices were distributed in eight US states.

    Product
    Consist of a distal connection mechanism (either standard Terumo Ball Quick Connect or Terumo Shaft Quick Connect), an articulating arm with links, a handle for providing cable tension, and a mechanism to mount the device to a sternal retractor.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0565-2021·2020-12-23

    Terumo Capiox FX25 Advance Oxygenator Recall for Configuration Error

    Terumo is recalling Capiox FX25 Advance Oxygenators because they may have the incorrect East or West configuration. The recall affects 5,079 units distributed in the US and abroad.

    Product
    (1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, Sterile Catalog Number: 3CX*FX25RWC (2) Capiox FX25 (East) Advance Oxygenator - FX25 with right port, 4-liter Reservoir, XC modified to Advance Catalog Number…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2960-2020·2020-09-23

    Terumo Ultrasonic Air Sensor Recall Due to Backwards Crimped Wire

    Terumo is recalling three Ultrasonic Air Sensors because a crimped wire was installed backwards, potentially causing false air bubble alarms during extracorporeal procedures.

    Product
    Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x1/16 UAS BLK, Catalog Number 5785 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2961-2020·2020-09-23

    Terumo Ultrasonic Air Sensor Recalled for Intermittent Connection Defect

    Terumo is recalling two Ultrasonic Air Sensors because a manufacturing defect may cause intermittent connections and false air bubble alarms during extracorporeal procedures.

    Product
    Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x3/32, Catalog Number 5791 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2959-2020·2020-09-23

    Terumo Ultrasonic Air Sensor Recall Due to Intermittent Connection

    Terumo is recalling 100 Ultrasonic Air Sensors because a manufacturing defect may cause false air bubble alarms during extracorporeal procedures.

    Product
    Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1108-2020·2020-02-19

    Terumo Advanced Perfusion System 1 Roller Pump Recall for Shaft Failure

    Terumo is recalling 33 Advanced Perfusion System 1 Roller Pumps because the pump shafts may fail, potentially causing the pump to stop during surgery.

    Product
    Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1111-2020·2020-02-19

    Terumo Advanced Perfusion Assembly Pump Guts Recalled for Potential Shaft Failure

    Terumo is recalling 11 Advanced Perfusion Assembly Pump Guts units because pump shafts may fail, potentially causing the pump to stop during surgery.

    Product
    Terumo Advanced Perfusion Assembly Pump Guts, 4 Inch, Catalog Number 801805 - Product Usage: is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1109-2020·2020-02-19

    Terumo Advanced Perfusion System 1 Roller Pump Recall for Shaft Failure

    Terumo is recalling 106 Advanced Perfusion System 1 Roller Pumps because the pump shafts may fail, potentially causing the pump to stop during surgery.

    Product
    Terumo Advanced Perfusion System 1 Roller Pump, 6 Inch, Catalog Number 816571 - Product Usage: is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1110-2020·2020-02-19

    Terumo Advanced Perfusion Assembly Pump Guts Recalled for Potential Shaft Failure

    Terumo is recalling 16 Advanced Perfusion Assembly Pump Guts units because the pump shafts may fail, potentially causing the pump to stop during surgery.

    Product
    Terumo Advanced Perfusion Assembly Pump Guts, 6 Inch, Catalog Number 801798 - is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2104-2019·2019-08-07

    Terumo Capiox NX19 Oxygenator Recalled for Non-Functioning Thermistor

    Terumo is recalling Capiox NX19 Oxygenators because they may contain a non-functioning arterial thermistor component.

    Product
    Terumo Capiox NX19 Oxygenator (West Orientation P/N:CXOT281 Exported only to Japan for inclusion into convenience kits that are sold only within the Japanese market
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2103-2019·2019-08-07

    Terumo Capiox NX19 Oxygenator Recalled for Non-Functioning Thermistor

    Terumo is recalling 78 Capiox NX19 Oxygenators exported to Japan and Belgium because they may contain a non-functioning arterial thermistor component.

    Product
    Terumo Capiox NX19 Oxygenator (East Orientation), P/N: CXOT280 Exported only to Japan for inclusion into convenience kits that are sold only within the Japanese market
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2102-2019·2019-08-07

    Terumo Capiox NX19 Oxygenator Recalled for Non-Functioning Thermistor

    Terumo is recalling 52 Capiox NX19 Oxygenators because they may contain a non-functioning arterial thermistor component.

    Product
    Terumo Capiox NX19 Oxygenator With Reservoir (West Orientation) P/N: 3CX*NX19RW
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2101-2019·2019-08-07

    Terumo Capiox NX19 Oxygenator Recalled for Non-Functioning Arterial Thermistor

    Terumo is recalling 16 Capiox NX19 Oxygenators with a non-functioning arterial thermistor component. The units were distributed in Michigan, Japan, and Belgium.

    Product
    Terumo Capiox NX19 Oxygenator With Reservoir (East Orientation) P/N: 3CX*NX19RE
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1769-2019·2019-06-19

    Terumo Electronic Patient Gas System Recalled for Control Slider Failure

    Terumo is recalling 72 Electronic Patient Gas Systems used in cardio-pulmonary bypass because a fault disables control sliders, forcing use of local knobs.

    Product
    Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas Blender, Catalog #: 801188, UDI 00886799000588. Used in cardio-pulmonary bypass.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1337-2019·2019-05-22

    Terumo Electronic Patient Gas System Recalled for Inaccurate Gas Flow Rate

    Terumo is recalling 1,258 Electronic Patient Gas Systems because a flowmeter defect may cause inaccurate gas flow rate output.

    Product
    Terumo Advanced Perfusion System 1 Electronic Patient Gas System, Catalog # 801188, UDI 00886799000588 The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit.
    Category
    Medical Device
    Distribution
    Distributed nationwide