Steris 4085 General Surgical Table Recalled for Pressure Switch Defect
Steris Corporation is recalling 3,689 STERIS 4085 General Surgical Tables because a defective pressure switch may cause the hand control to incorrectly indicate that floor lock feet are locked.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product where injury has not yet been reported, which corresponds to a High severity score of 3 according to the rubric.
Plain-English summary
Steris Corporation is recalling 3,689 units of the STERIS 4085 General Surgical Table. This electro-hydraulically operated surgical table is designed to support all general surgical procedures. The recall covers all model numbers and includes units distributed worldwide, including across the United States and in various international markets.
The recall was initiated after Steris engineering analysis determined that the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure levels. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.
This defect poses a risk of harm to patients and medical staff if the table is used while the floor lock feet are not securely engaged. The source text does not report any specific injuries or illnesses associated with this defect. Users of these surgical tables should contact Steris Corporation for instructions on how to resolve the issue.
The recalled product
- Product
- The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical procedures.
- Manufacturer
- Steris Corporation
- Category
- Medical Device — Surgical Table
- Affected units
- 3,689
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All Model Numbers
Distribution
Related recalls
Same manufacturer · Steris Corporation
See all →- ModerateHarmonyAIR A-Series surgical light heads may show paint delamination
FDA (Devices) · 2026-06-24
- ModerateHarmonyAIR A-Series surgical light intensity control buttons may detach
FDA (Devices) · 2026-06-24
- ModerateALYON surgical light head intensity control buttons may detach
FDA (Devices) · 2026-06-24
- ModerateDisplay adapters recalled after the wrong configuration file was used
FDA (Devices) · 2026-02-18
- CriticalFrozen whole surti papdi recalled after testing positive for Salmonella
FDA (Devices) · 2025-12-17
Same hazard · mechanical failure
See all →- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighForest River Transit Buses Recalled for Wheelchair Lift Control Failure
NHTSA · 2026-09-01
- HighBlue Bird Recalls Vision School Buses Due to Wheelchair Lift Defect
NHTSA · 2026-08-31
- SevereMagic Tilt Watercraft Trailers Recalled Due to Winch Weld Failure
NHTSA · 2026-08-28