The Recall Desk
HighFDA (Devices)·Z-1906-2020·Announced 2020-05-13

Intuitive Surgical Recalls da Vinci SP Instrument Arm Drapes

Intuitive Surgical is recalling da Vinci SP Instrument Arm Drapes because they may tear, compromising the sterile barrier.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (sterile barrier breach) where injury has not yet been reported, fitting the High severity criteria.

Plain-English summary

Intuitive Surgical, Inc. is recalling 1,818 pouches of da Vinci SP Instrument Arm Drape (REF 430013-12). The recall was initiated because the instrument drape may tear while draping the Patient Side Cart, which presents a breach in the sterile barrier.

The affected product is distributed in the United States and South Korea. The source text does not specify the number of reported injuries or illnesses associated with this defect.

Healthcare facilities should stop using the affected drapes and contact Intuitive Surgical for further instructions.

The recalled product

Product
da Vinci SP Instrument Arm Drape. REF 430013-12 QTY 1 Pouch
Affected units
1,818

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Model Number: 430013-12 UDI Number: 00886874113486 Lot Numbers: All Lots

Distribution

Distributed in 17 states: