Siemens Artis zee Ceiling X-Ray System Recall for Loose Screws
Siemens is recalling 276 Artis zee ceiling-mounted X-ray systems because screws joining the pivot bearing to the ceiling support may work loose or snap off, posing a risk of severe injury.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The source text explicitly states that the defect 'can result in severe injuries to patients and operating personnel,' which aligns with the rubric criterion for 'significant injury/property-damage reports' or risk of severe harm, warranting a score of 4 (Severe).
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 276 units of the Siemens Artis zee ceiling-mounted Interventional Fluoroscopic X-Ray System (Model Number: 10094137). The recall also applies to Artis Q, Artis Pheno, or Artis Icono systems equipped with a DCS Extended or DCS Movable.
The recall was issued because individual screws in the connection joining the pivot bearing to the ceiling-mounted support may work loose or snap off. This defect can result in severe injuries to patients and operating personnel. The affected units were distributed nationwide.
Healthcare facilities should identify affected units by checking the serial numbers listed in the official recall notice. Siemens is the recalling firm and is responsible for addressing the defect.
The recalled product
- Product
- Siemens Artis zee ceiling -Interventional Fluoroscopic X-Ray System Model Number: 10094137
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 276
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 12 recalled by Siemens Medical Solutions USA, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 12 products in this recall →Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · structural failure
See all →- SevereMondraker Recalls Mountain Bicycles with ONOFF S9 Carbon Handlebars
CPSC · 2026-09-03
- HighFord F-650 SD Recall: Liftgate Platform Deck Pins May Break
NHTSA · 2026-08-20
- HighStryker Hoffmann II Connecting Rod Micro May Crack After Cleaning and Sterilization
FDA (Devices) · 2026-08-05
- HighStryker Hoffmann II Connecting Rod Micro May Crack After Cleaning and Sterilization
FDA (Devices) · 2026-08-05
- High