ConMed Recalls Hip Preservation System PreBent Burs Due to Metal Shavings
ConMed is recalling 991 Hip Preservation System PreBent Burs because an assembly error caused metal shavings to be released during use.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a risk of harm (metal shavings) where injury has not yet been reported, which aligns with the 'High' severity criteria in the rubric.
Plain-English summary
ConMed Corporation is recalling 991 pieces of the Hip Preservation System Signature Series PreBent Burs. The affected products include the PreBent Spherical Bur (Catalog Number HPS-HB11), PreBent Polishing Bur (Catalog Number HPS-HB12), and PreBent Oval Bur (Catalog Number HPS-HB13). These medical devices are intended for the resection of osseous tissue in arthroscopic and endoscopic surgical procedures.
The recall was initiated because metal shavings are released from the burs during use. This issue is caused by contact between the bur and the hood of the PreBent Bur assembly. The root cause is an assembly error during the manufacturing process that resulted in the inadvertent mixing of two sizes of bearings.
The affected units were distributed nationwide to California, Colorado, Florida, Indiana, Maryland, Minnesota, North Carolina, New York, Ohio, Texas, Utah, and Virginia, as well as internationally to Canada and Wales. Specific lot numbers are listed in the structured product information. No illnesses or injuries are reported in the source text.
The recalled product
- Product
- Hip Preservation System Signature Series PreBent Burs under the following labels: 1) PreBent Spherical Bur, 5.5mm x 19cm Catalog Number HPS-HB11, 2) PreBent Polishing Bur, 5.5mm x 19cm Catalog Number HPS-HB12 and 3) PreBent Oval Bur, 6.0mm x 19cm Catalog Number HPS-HB13 The…
- Manufacturer
- ConMed Corporation
- Affected units
- 991
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Lot Numbers: 1) 569449
- 577583
- 577969
- 579655
- 580365
- 581176
- 581198
- 582216
- 587320
- 594569
- 658219
- 669615
- 673913 & 688136
- 2) 569462
- 577800
- 577971 & 579652
- 3) 569468
- 577835
- 577974
- 579650
Distribution
Related recalls
Same manufacturer · ConMed Corporation
See all →- ModerateConMed Suture Anchor Recalled for Incorrect Patient Implant Card
FDA (Devices) · 2026-09-02
- HighConMed AirSeal 12mm blunt tip 100mm ports may trigger overpressure alerts
FDA (Devices) · 2026-06-24
- HighConMed AirSeal 12mm bladeless optical obturators may trigger overpressure alerts
FDA (Devices) · 2026-06-24
- HighConMed AirSeal 5mm 100mm optical obturator ports may trigger overpressure alerts
FDA (Devices) · 2026-06-24
- HighConMed AirSeal 8mm 100mm optical obturator ports may trigger overpressure alerts
FDA (Devices) · 2026-06-24
Same hazard · metal shavings
See all →- HighDQ Chocolate Reduced Fat Ice Cream Mix Recalled Due to Metal Shavings
FDA (Food) · 2026-08-05
- HighDURAFLOW 2 Vascular Access Device Recalled for Loose Metal Shavings
FDA (Devices) · 2024-05-15
- HighHemodialysis vascular access kit recalled for loose metal shavings
FDA (Devices) · 2024-05-15
- HighVascular access device recalled due to loose metal shavings
FDA (Devices) · 2024-05-15
- HighHemodialysis vascular access kit recalled due to metal shavings risk
FDA (Devices) · 2024-05-15
Similar recalls
- ModerateCompress Device Short Anchor Plugs recalled due to metal burrs
FDA (Devices) · 2023-11-29
- HighBiomet Compress Device Anchor Plugs recalled for metal burrs in transverse holes
FDA (Devices) · 2023-11-29
- HighCompress Device Short Anchor Plugs Recalled for Metal Burrs
FDA (Devices) · 2023-11-29
- ModerateOrthopedic Anchor Plugs Recalled for Metal Burrs Preventing Proper Installation
FDA (Devices) · 2023-11-29
- HighBiomet Compress Device Short Anchor Plug Recalled Due to Metal Burrs
FDA (Devices) · 2023-11-29
- ModerateCompress Device Short Anchor Plug recalled for potential metal burrs
FDA (Devices) · 2023-11-29