ConMed AirSeal 12mm blunt tip 100mm ports may trigger overpressure alerts
ConMed AirSeal 12 mm Access Ports and Obturators with Blunt Tip, 100 mm length, are being recalled because of the potential for overpressure alerts. 1,261 packs affected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall for a malfunction arising during laparoscopic use, a risk of harm with no injuries reported in the source.
Plain-English summary
The ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 100 mm length, catalog number iASB12-100, six units per pack, is being recalled. The product is a disposable endoscopic trocar and cannula used with a carbon dioxide insufflator for laparoscopy. The action covers 1,261 packs, totalling 7,566 units.
The stated reason is the potential for overpressure alerts. The FDA classified the action as Class II.
Affected product is UDI-DI 10851885004310, spanning the lot numbers listed in the recall notice including 202510034 and 202505134. Distribution was nationwide in the United States and internationally to Australia, Austria, Belgium and other markets.
Facilities should identify affected stock using the catalog number, UDI-DI and lot numbers in the recall notice, quarantine it, and follow the instructions issued by ConMed.
The recalled product
- Product
- ConMed AirSeal 12mm Access Port and Obturator with Blunt Tip, 100mm Length. Catalog Number: iASB12-100. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.
- Manufacturer
- ConMed Corporation
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (14)
- Catalog Number: iASB12-100. UDI-DI: 10851885004310. Lot Numbers: 202510034
- 202505134
- 202507234
- 202510094
- 202503184
- 202408306
- 202407044
- 202412194
- 202512014
- 202512024
- 202410085
- 202411264. *OUS added 06/10/2026 expansion: 202510034
- 202411264
- 202510094. Manufactured from 14 June 2024 to 12 January 2026.
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 11 recalled by ConMed Corporation under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 11 products in this recall →Related recalls
Same manufacturer · ConMed Corporation
See all →- ModerateConMed Suture Anchor Recalled for Incorrect Patient Implant Card
FDA (Devices) · 2026-09-02
- HighConMed AirSeal 5mm 100mm optical obturator ports may trigger overpressure alerts
FDA (Devices) · 2026-06-24
- HighConMed AirSeal 8mm 100mm optical obturator ports may trigger overpressure alerts
FDA (Devices) · 2026-06-24
- HighConMed AirSeal 12mm bladeless optical obturators may trigger overpressure alerts
FDA (Devices) · 2026-06-24
- HighConMed AirSeal 12mm palm grip 150mm ports may trigger overpressure alerts
FDA (Devices) · 2026-06-24
Same hazard · device malfunction
See all →- SevereMedtronic MiniMed Flex Insulin Pumps Recalled for Failed Verification Tests
FDA (Devices) · 2026-10-14
- ModerateWilson-Cook Recalls Hemospray Endoscopic Hemostat Due to Adherence Issues
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Et Control Software Recalled for Anesthetic Delivery Interruption
FDA (Devices) · 2026-09-23
Similar recalls
- HighConMed Anchor Tissue Retrieval System Recalled for Seal Defects
FDA (Devices) · 2019-05-15
- ModerateBovie Medical Recalls ConMed Disposable Bipolar Ablators Due to Packaging Defect
FDA (Devices) · 2021-03-10
- HighNorthgate Recalls ConMed Abdominal Insufflators Due to Incorrect Tubing Material
FDA (Devices) · 2021-03-03
- CriticalAdult Manual Resuscitator with Medium Adult Mask recalled nationwide
FDA (Devices) · 2025-06-11
- HighConMed Anchor Tissue Retrieval System Recalled for Sterility Seal Defects
FDA (Devices) · 2019-05-15
- ModerateLaparoscopic trocars recalled over a potentially extended blade obturator length
FDA (Devices) · 2026-01-14