The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1825-2024·Announced 2024-05-22

Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit Recall Due to CSF Leakage Risk

Sophysa is recalling the Pressio Ventricular Intracranial Pressure Monitoring Kit due to manufacturing defects causing cerebrospinal fluid leakage from the catheter. Undetected leaks during implantation could result in infection risk to patients.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: FDA Class II recall for a manufacturing defect creating risk-of-harm through potential CSF leakage and subsequent infection. No specific illnesses or injuries are documented in the source text, placing this at High severity per the rubric criterion for risk-of-harm products where injury has not yet been reported.

Plain-English summary

The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit (PSO-VT) is a medical device intended for monitoring intracranial ventricular pressure and cerebrospinal fluid drainage applications.

An increase in claims related to cerebrospinal fluid (CSF) leakage from the proximal end of the catheter has been reported. The manufacturer has identified the cause as a manufacturing error involving perforation of the inner channel of the catheter during stylet insertion. If a leak goes undetected during implantation, it could pose an infectious risk to the patient.

The recall affects 3,487 units distributed outside the United States to the following locations: Arizona, New York, and Texas within the United States, and internationally to Argentina, China, Czech Republic, France, Greece, Hungary, Japan, Pakistan, Portugal, Romania, South Africa, Sweden, and Vietnam.

Patients or healthcare providers with questions about this device should contact Sophysa S.A. or their healthcare provider for guidance on device management.

The recalled product

Product
The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application.…
Manufacturer
Sophysa S.A.
Affected units
3,487

Distribution

Distributed nationwide across the United States.

The notice also names 3 states specifically:

Part of a larger recall action

This product is one of 2 recalled by Sophysa S.A. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.