FDA (Devices) recall action 94464

Sophysa S.A. recalled 2 products on 2024-05-22

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-05-22
Critical
0
Units
3,487

Why these products were recalled

An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-05-22

    Sophysa Pressio Ventricular ICP Monitoring Kit Recalled for Manufacturing Defect Causing CSF Leakage

    Sophysa is recalling the Pressio Ventricular Intracranial Pressure Monitoring Kit due to a manufacturing defect that causes cerebrospinal fluid leakage. An undetected leak during implantation could pose an infectious risk to patients.

    Product
    The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature with tunneling (PSO-VTT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2024-05-22

    Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit Recall Due to CSF Leakage Risk

    Sophysa is recalling the Pressio Ventricular Intracranial Pressure Monitoring Kit due to manufacturing defects causing cerebrospinal fluid leakage from the catheter. Undetected leaks during implantation could result in infection risk to patients.

    Product
    The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application.…
    Category
    Distribution
    3 states