The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1825-2021·Announced 2021-06-16

Cervical Plate System screw locking mechanism may fail during or after surgery

Alphatec Insignia Anterior Cervical Plate System screw blocking mechanism may disassociate during surgery or unlock afterward. No injuries reported.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Class II FDA medical device recall where no injuries have been reported and the hazard is characterized as potential. Per the rubric, risk-of-harm products where injury has not yet been reported receive a High severity rating.

Plain-English summary

The Atec Insignia Anterior Cervical Plate System (REF 136-0474, 4-level, 74 mm) manufactured by Alphatec Spine, Inc., is being recalled due to a potential defect in the screw blocking mechanism. The mechanism may disassociate during or after surgery, or may unlock after surgery.

The recall affects 5 units distributed across the United States, including Florida, California, Illinois, Oklahoma, Texas, Connecticut, Indiana, North Carolina, Washington, Hawaii, Idaho, Massachusetts, South Carolina, New Mexico, and New York. The specific lot code is 8609333R.

Patients who have received this implant should contact their surgeon or healthcare provider immediately to discuss potential risks and appropriate follow-up or monitoring. Healthcare providers should verify their inventory for this specific implant and assess patient status if it was used.

The recalled product

Product
Atec Insignia Anterior Cervical Plate System, REF 136-0474 Insignia, ACP, 4-Level, 74 mm, Rx Only, Non-Sterile, UDI: (01)00190376268699
Affected units
5

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • 8609333R

Distribution

Distributed nationwide across the United States.

The notice also names 15 states specifically:

Part of a larger recall action

This product is one of 29 recalled by Alphatec Spine, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 29 products in this recall →

Similar recalls