The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-2273-2012·Announced 2012-09-05

Alphatec Recalls Trestle Luxe Anterior Cervical Plating System Components

Alphatec Spine is recalling 38 units of the Trestle Luxe Anterior Cervical Plating System because the Slide Alignment Tool does not fit the plate screws.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, but the source text explicitly states no illnesses or injuries were reported, and the defect is a mechanical fit issue rather than a high-risk pathogen or structural failure causing immediate harm.

Plain-English summary

Alphatec Spine, Inc. is recalling 38 units of the Trestle Luxe Anterior Cervical Plating System, Product code KWQ. The recall involves specific lot numbers: 6629701, 6594701, 6434301, 6434302, and 6434302R1. The product was distributed nationwide.

The recall was initiated after Alphatec Spine received a complaint on May 9, 2012, reporting that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721) does not properly fit into the hexalobe screw on the Trestle Luxe Plate. The Trestle Luxe system is a temporary device used to stabilize the cervical spine during bone fusion development.

Healthcare providers and patients who have used or are using this specific device should contact Alphatec Spine for further instructions. The source text does not report any illnesses or injuries associated with this defect.

The recalled product

Product
TRESTLE LUXE¿ Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and…
Affected units
38

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (5)

  • Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701
  • 6594701
  • 6434301
  • 6434302
  • 6434302R1.

Distribution

Distributed nationwide across the United States.

Similar recalls