Alphatec Recalls Trestle Luxe Anterior Cervical Plating System Components
Alphatec Spine is recalling 38 units of the Trestle Luxe Anterior Cervical Plating System because the Slide Alignment Tool does not fit the plate screws.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text explicitly states no illnesses or injuries were reported, and the defect is a mechanical fit issue rather than a high-risk pathogen or structural failure causing immediate harm.
Plain-English summary
Alphatec Spine, Inc. is recalling 38 units of the Trestle Luxe Anterior Cervical Plating System, Product code KWQ. The recall involves specific lot numbers: 6629701, 6594701, 6434301, 6434302, and 6434302R1. The product was distributed nationwide.
The recall was initiated after Alphatec Spine received a complaint on May 9, 2012, reporting that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721) does not properly fit into the hexalobe screw on the Trestle Luxe Plate. The Trestle Luxe system is a temporary device used to stabilize the cervical spine during bone fusion development.
Healthcare providers and patients who have used or are using this specific device should contact Alphatec Spine for further instructions. The source text does not report any illnesses or injuries associated with this defect.
The recalled product
- Product
- TRESTLE LUXE¿ Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bon
- Manufacturer
- Alphatec Spine, Inc.
- Hazard
- Affected units
- 38
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701
- 6594701
- 6434301
- 6434302
- 6434302R1.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Alphatec Spine, Inc.
See all →- HighSafeOp Patient Interface Neuromonitoring Device Recalled Due to Software Issue
FDA (Devices) · 2026-07-29
- ModerateNavigation osteotomes recalled over incorrect array connection geometry
FDA (Devices) · 2026-02-18
- ModerateSpinal navigation instruments recalled over incorrect array connection geometry
FDA (Devices) · 2026-02-18
- HighAlphatec Calibrate CCX Interbody Implants Recalled for Post-Operative Collapse
FDA (Devices) · 2025-02-05
- HighIdentiTi ALIF surgical graft bolts recalled due to intraoperative breakage
FDA (Devices) · 2024-03-06
Same hazard · mechanical defect
See all →- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- ModerateSurgical driver systems recalled over out of the box driver wobble
FDA (Devices) · 2025-09-03
- HighBeckman Coulter DxI 9000 Analyzer wash wheel mixer defect may prevent operation
FDA (Devices) · 2025-03-19
- SevereBaxter Spectrum IQ Infusion System may have missing motor mounting screws
FDA (Devices) · 2025-03-12
- HighVNS Therapy Generator May Stop Delivering Stimulation Due to Mechanical Defect
FDA (Devices) · 2025-02-12