MiniMed 620G insulin pump delivery varies with pump height changes
Medtronic is recalling MiniMed 620G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Rubric level 3 covers risk-of-harm products where injury has not yet been reported: the source lists death among the potential consequences of the mechanism but records no reported death, injury or hospitalisation.
Plain-English summary
Medtronic MiniMed, Inc. is recalling the MiniMed 620G Insulin Pump, model MMT-1750. About 799 units were distributed across the United States nationwide and more than eighty other countries. The FDA has classified this recall as Class II.
All Medtronic MiniMed infusion pumps, covering the Paradigm series, 600 series and BLE 700 series, were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Medtronic states that over-delivery, occurring when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma or death, and that under-delivery, occurring at lowered pump height, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis or death.
Affected pumps are listed by UDI-DI and user guide part version in the FDA enforcement record.
Users and clinicians can identify affected pumps by the model number above, and may contact Medtronic MiniMed, Inc. for further information.
The recalled product
- Product
- MiniMed 620G Insulin Pump (MMT-1750)
- Manufacturer
- Medtronic MiniMed, Inc.
- Category
- Medical Device — Insulin pump
- Affected units
- 799
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (17)
- 643169721319
- 763000375461
- 763000253189: Instr
- Startup Guide
- 620G 4.10
- EN(6025951-022 1)
- INSTR.
- NGP 620G
- BOLUS WIZARD
- EN(6026006-023 1)
- Instr.
- 620G
- Sftwr. 4.10
- EN.(6026006-024 1)
- INSTRUCTIONS MMT-746EN PUMP 1510/1710 EN(7005933-023 1)
- DOC PKG MMT-746EN NGP 1710 EN(7005933-024 1)
- Startup Guide Assembly MMT-755EN(620G(M982438A021 1).
Distribution
Part of a larger recall action
This product is one of 14 recalled by Medtronic MiniMed, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 14 products in this recall →Related recalls
Same manufacturer · Medtronic MiniMed, Inc.
See all →- HighMedtronic MiniMed 780G Insulin Pumps Recalled for Chemical Damage Risk
FDA (Devices) · 2026-08-26
- SevereMedtronic MiniMed 670G Insulin Pumps Recalled Due to Chemical Damage Risk
FDA (Devices) · 2026-08-26
- HighMedtronic MiniMed 640G Insulin Pumps Recalled for Chemical Damage Risk
FDA (Devices) · 2026-08-26
- HighMiniMed Go App Recalled Due to Missed Insulin Alerts
FDA (Devices) · 2026-07-29
- HighMiniMed 780G Insulin Pump Software Anomaly Recall
FDA (Devices) · 2026-07-15
Same hazard · insulin delivery error
See all →- HighMiniMed 770G insulin pump delivery varies with pump height changes
FDA (Devices) · 2026-04-08
- HighMiniMed 720G insulin pump delivery varies with pump height changes
FDA (Devices) · 2026-04-08
- HighParadigm insulin pump delivery varies with pump height changes
FDA (Devices) · 2026-04-08
- HighMiniMed 740G insulin pump delivery varies with pump height changes
FDA (Devices) · 2026-04-08
- HighMiniMed 780G insulin pump delivery varies with pump height changes
FDA (Devices) · 2026-04-08