Philips Brilliance CT Big Bore Recalled for Radio Frequency Emission Tolerance Issues

Philips is recalling 454 Brilliance CT Big Bore systems because internal testing found radio frequency emissions exceeded IEC 60601-1-2 standards.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, and the source text does not report any illnesses, injuries, or hospitalizations, fitting the criteria for a moderate severity score.

Plain-English summary

Philips Medical Systems (Cleveland) Inc. is recalling 454 units of the Brilliance CT Big Bore Oncology and Radiology Whole Body Computed Tomography X-Ray Systems. The recall was initiated after internal testing revealed that the devices were out of tolerance for radio frequency emissions.

Specifically, testing at the 48MHz frequency indicated that the Brilliance CT Big Bore emitted signals 3.5dB uV/meter higher than the applicable IEC 60601-1-2 standard specification. The affected units were distributed worldwide, including in the United States and numerous international countries.

No specific consumer actions are detailed in the source text beyond the identification of the defect. The recall is classified as Class II by the FDA.

The recalled product

Product
Brilliance CT Big Bore Oncology Whole Body Computed Tomography X-Ray System & Brilliance CT Big Bore Radiology Whole Body Computed Tomography X-Ray System. Product Usage: The Brilliance CT Big Bore is a Whole Body Computed Tomography X-Ray System which produces cross-secti
Affected units
454

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Brilliance CT Big Bore Oncology: Model #728243
  • Serial #s: 7185
  • 7187-7189
  • 7191-7199
  • 7201
  • -7243
  • 7245-7251
  • 7253-7267
  • 7269
  • 7270
  • 7272-7297
  • 7299
  • 7300
  • 7302
  • 7304-7317
  • 7319-7374
  • 7376-7379
  • 7381
  • 7382
  • 7385

Distribution

Distributed nationwide across the United States.

The notice also names 44 states specifically:

Same hazard · rf emission