Philips Brilliance 64 CT System Software Defect Recall
Philips Healthcare is recalling 423 Brilliance 64 CT systems in the US due to software defects causing incorrect scan lengths and dose calculations.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, which typically corresponds to a High severity score (3) when no specific hospitalization reports or Class I classification are present in the source text.
Plain-English summary
Philips Healthcare is recalling 423 units of the Brilliance 64 Computed Tomography X-Ray System in the United States. The recall is due to software defects in versions 4.13, 4.14, and 4.15 that affect image reconstruction and dose calculation.
The identified software defects result in several issues, including shortened sagittal results for axial scans, incorrect display of single series during multi-planner reconstruction, and changes in scan length when the field of view is adjusted. Additionally, the DoseRight algorithm may miscalculate, leading to incorrect CTDI values, and scan lengths may change during subsequent axial results or near specific lengths of 135mm or 184mm.
The affected systems are distributed across 31 states and the District of Columbia, including Arizona, Pennsylvania, Florida, and California. No specific instructions for consumers are provided in the source text, but the recall is classified as FDA Class II.
The recalled product
- Product
- Philips Healthcare Brilliance 64 Computed Tomography X-Ray System
- Manufacturer
- Philips Healthcare
- Affected units
- 423
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Software v. 4.13
- 4.14 and 4.15.
Distribution
Part of a larger recall action
This product is one of 7 recalled by Philips Healthcare under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 7 products in this recall →Related recalls
Same manufacturer · Philips Healthcare
See all →- HighPhilips Incisive CT Software Version 5.0 Recall for Scan Parameter Issues
FDA (Devices) · 2023-10-11
- HighPhilips StentBoost Live imaging system acquisition malfunction recall
FDA (Devices) · 2022-04-13
- HighAzurion Medical Imaging System Software Defect Affects Patient Dose Control
FDA (Devices) · 2022-03-30
- HighPhilips CombiDiagnost R90 X-ray systems missing required radiation warning labels
FDA (Devices) · 2021-12-01
- ModerateMedical X-Ray Systems Missing Required Radiation Warning Labels
FDA (Devices) · 2021-12-01
Same hazard · software defect
See all →- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25