Brainlab ExacTrac 5.5 Image Processing System Recalled for Plan Blending Error
Brainlab AG is recalling ExacTrac 5.5 image processing systems because a specific workflow may generate unintended blends of separate patient treatment plans.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device defect that could lead to incorrect treatment planning, representing a risk of harm where injury has not yet been reported.
Plain-English summary
Brainlab AG is recalling 40 ExacTrac 5.5 Image Processing Systems used for patient positioning in radiation therapy or radiosurgery. The recall involves software version 5.5.5 and specific upgrade kits.
The recall is due to a software defect where using a specific plan update workflow with a non-Brainlab treatment planning system may generate an unintended blend of two initially separate plans for the same patient. This could result in incorrect treatment planning data.
The affected systems are distributed worldwide, including in the United States (AZ, AR, CA, FL, KY, MA, NE, NY, TN, TX) and in Argentina, Australia, Brazil, Finland, India, Japan, Netherlands, Oman, Russia, and Taiwan. Healthcare facilities should contact Brainlab AG for instructions on how to resolve the issue.
The recalled product
- Product
- ExacTrac 5.5, an Image Processing System used for patient positioning for radiation therapy or radiosurgery.
- Manufacturer
- Brainlab AG
- Affected units
- 40
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (14)
- 2) 20820 ET SOFTWARE UPDATE 5.0 TO 5.5
- 3) 49905D EXACTRAC COMPLETE SYSTEM UPG 3.X TO 5.5
- 4) 49908C EXACTRAC SOFTWARE UPGRADE 3.X TO 5.5
- 5) 49909A EXACTRAC SOFTWARE UPGRADE 4.X TO 5.5
- 6) 49918 ET SOFTWARE UPDATE 5.5.X TO 5.5.5
- 7) 49919 ET SOFTWARE UPDATE 5.0.X TO 5.5.5
- 8) 49660A ET INFRARED SOFTWARE 5.5.5
- 9) 49961 ET DATA PREP/REVIEW SYSTEM
- 10) 49986 ET SYSTEM UPGRADE 3.X TO 5.5 NEW VAULT
- 11) 49987 ET SYSTEM UPGRADE 4.X TO 5.5 NEW VAULT
- 12) 49988 ET SYSTEM UPGRADE 4.X TO 5.5 SAME VAULT
- 13) 49989 ET SYSTEM UPGRADE 5.0 TO 5.5 NEW VAULT
- 14) 49990 ET SYSTEM UPGRADE 5.0 TO 5.5 SAME VAULT
- 15) 49991 ET SYSTEM UPGRADE 5.5 TO 5.5 NEW VAULT
Distribution
Related recalls
Same manufacturer · Brainlab AG
See all →- HighMedical Device Software May Merge Patient Records Without Notification
FDA (Devices) · 2025-03-26
- HighSurgical navigation system recalled for potential mechanical instability
FDA (Devices) · 2023-04-05
- HighRadiation Therapy Software May Not Properly Execute Heart-Protection Breath Hold
FDA (Devices) · 2023-03-22
- HighExacTrac Dynamic Software Yaw Angle Error May Affect Radiotherapy Positioning
FDA (Devices) · 2022-03-16
- HighRadiotherapy positioning software may delay patient movement display
FDA (Devices) · 2021-06-16
Same hazard · software defect
See all →- HighPhilips Spectral CT Systems Recalled for Software Issues Affecting Safety
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- ModerateUnited Imaging uCT 550 CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uMI Panvivo S PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
Similar recalls
- HighBrainlab ExacTrac Software Recall for Potential Radiation Misalignment
FDA (Devices) · 2014-08-13
- SevereBrainlab ExacTrac 6.0 Radiation Positioning System Recalled for Targeting Error
FDA (Devices) · 2014-05-07
- SevereBrainlab ExacTrac Dynamic Radiation Therapy System Software Recall
FDA (Devices) · 2021-03-17
- HighBrainlab ExacTrac 6.0 Patient Positioning System Recalled for Display Error
FDA (Devices) · 2015-09-02
- HighBrainlab ExacTrac 6.x Patient Positioning Systems Recalled for Positioning Errors
FDA (Devices) · 2015-04-01
- HighBrainlab ExacTrac Software Recall for Beam Authorization Misinterpretation Risk
FDA (Devices) · 2018-03-21