Siemens Dimension Vista Triglyceride Assay Recalled for Interference
Siemens is recalling Dimension Vista Triglyceride assays because N-Acetylcysteine and Metamizole can cause falsely depressed results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where a defect (interference) can lead to incorrect clinical results (falsely depressed triglycerides), posing a risk of harm to patients, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling Dimension Vista Assays: Triglycerides (TRIG) (Catalog # DF69A). The recall involves 137,422 units distributed nationwide to various states and territories.
The recall was initiated because Siemens became aware of interference from N-Acetylcysteine (NAC) and Metamizole (Dipyrone) with Trinder and Trinder-like reaction assays. The company confirmed that falsely depressed results may occur on samples drawn from patients receiving these specific medications. NAC is noted as the accepted antidote for acetaminophen toxicity.
Healthcare facilities and laboratories should stop using the affected assays and contact Siemens for further instructions regarding the recall.
The recalled product
- Product
- Dimension Vista Assays: Triglycerides (TRIG)
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 137,422
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Device Listing: D011436 - Triglycerides (TGL)
- Catalog # DF69A
Distribution
Part of a larger recall action
This product is one of 16 recalled by Siemens Healthcare Diagnostics, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 16 products in this recall →Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · interference
See all →- ModerateOrtho Clinical Diagnostics Recalls VITROS XT Chemistry Products ALB-TP Slides
FDA (Devices) · 2021-01-20
- HighBeckman Coulter Total Bilirubin Reagent Recalled for Acetaminophen Interference
FDA (Devices) · 2020-03-11
- HighRoche Tina-quant sTfR Assay Recalled for Rheumatoid Factor Interference
FDA (Devices) · 2019-10-16
- ModerateRoche Tina-quant sTfR Assay Recalled for Rheumatoid Factor Interference
FDA (Devices) · 2019-10-16
- ModerateTosoh AIA-PACK B12 Recalled Due to Potential Fluorescein Interference
FDA (Devices) · 2019-01-02
Similar recalls
- ModerateSiemens Dimension Vista Glucose Assay Recalled for Drug Interference
FDA (Devices) · 2018-05-16
- ModerateSiemens Dimension Vista AST Assay Recalled for Drug Interference
FDA (Devices) · 2018-05-16
- ModerateSiemens Dimension Vista Thyroxine Assay Recalled for Drug Interference
FDA (Devices) · 2018-05-16
- HighSiemens Dimension Vista Ammonia Assay Recalled for Drug Interference
FDA (Devices) · 2018-05-16
- ModerateSiemens Dimension Thyroxine Assay Recalled for Drug Interference
FDA (Devices) · 2018-05-16
- ModerateSiemens Dimension AST Assay Recalled for Drug Interference
FDA (Devices) · 2018-05-16