The Recall Desk

FDA (Devices) recall action 73667

Siemens Healthcare Diagnostics, Inc. recalled 16 products on 2016-05-18

The FDA (Devices) publishes a recall like this one as 16 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 16 pages.

Products
16
Announced
2016-05-18
Critical
0
Units
2,198,752

Why these products were recalled

Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–16 of 16

  • ModerateFDA (Devices)··2016-05-18

    Siemens ADVIA Chemistry Triglycerides Reagent Recalled for Interference

    Siemens Healthcare Diagnostics is recalling ADVIA Chemistry Triglycerides reagents because N-Acetylcysteine and Metamizole can cause falsely depressed test results.

    Product
    ADVIA CHEMISTRY TRIGLYCERIDES_2 (TRIG_2) REAGENT & ADVIA CHEMISTRY TRIGLYCERIDES_2 Concentrated (TRIG_c) REAGENT
    Category
    Distribution
    52 states
  • ModerateFDA (Devices)··2016-05-18

    Siemens Direct HDL Cholesterol Assay Recalled for Interference

    Siemens is recalling Direct HDL Cholesterol assays because N-Acetylcysteine and Metamizole can cause falsely depressed results.

    Product
    Dimension Assays: Direct HDL Cholesterol (AHDL)
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2016-05-18

    Siemens Recalls Direct HDL Cholesterol Assay Due to Interference

    Siemens Healthcare Diagnostics is recalling the Advia Chemistry Assay: Direct HDL Cholesterol due to potential falsely depressed results caused by NAC or Metamizole interference.

    Product
    Advia Chemistry Assay: Direct HDL Cholesterol (D-HDL)
    Category
    Distribution
    52 states
  • ModerateFDA (Devices)··2016-05-18

    Siemens LDL Cholesterol Assay Recalled for Interference

    Siemens is recalling Advia Chemistry Assay: LDL Cholesterol, Direct because N-Acetylcysteine and Metamizole can cause falsely depressed results.

    Product
    Advia Chemistry Assay: LDL Cholesterol, Direct
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2016-05-18

    Siemens Uric Acid Assay Recalled Due to Drug Interference

    Siemens Healthcare Diagnostics is recalling Dimension Uric Acid assays because N-Acetylcysteine and Metamizole can cause falsely low results.

    Product
    Dimension Assays: Uric Acid (URCA)
    Category
    Distribution
    52 states
  • SevereFDA (Devices)··2016-05-18

    Siemens Uric Acid Assays Recalled Due to Drug Interference

    Siemens Healthcare Diagnostics is recalling Dimension Vista Uric Acid assays because NAC and Metamizole can cause falsely depressed results.

    Product
    Dimension Vista Assays: Uric Acid (URCA)
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2016-05-18

    Siemens ADVIA Chemistry Glucose Oxidase Assay Recalled for Interference

    Siemens is recalling ADVIA Chemistry Glucose Oxidase Assays because N-Acetylcysteine and Metamizole can cause falsely depressed results.

    Product
    ADVIA CHEMISTRY GLUCOSE OXIDASE ASSAY AND ADVIA Chemistry Glucose & Concentrated Glucose Oxidase Assay and ADVIA Chemistry Glucose.
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2016-05-18

    Siemens Direct HDL Cholesterol Assay Recalled for Interference

    Siemens Healthcare Diagnostics is recalling Direct HDL Cholesterol assays because N-Acetylcysteine and Metamizole can cause falsely depressed results.

    Product
    Dimension Vista Assays: Direct HDL Cholesterol (AHDL)
    Category
    Distribution
    52 states
  • ModerateFDA (Devices)··2016-05-18

    Siemens Uric Acid Assay Recalled Due to Interference

    Siemens Healthcare Diagnostics is recalling Uric Acid assays because N-Acetylcysteine and Metamizole can cause falsely depressed results.

    Product
    Advia Chemistry Assay: Uric Acid and Uric Acid, concentrated
    Category
    Distribution
    52 states