Siemens Dimension Glucose Assay Recalled for Drug Interference
Siemens Healthcare Diagnostics is recalling Dimension Glucose assays because they may produce inaccurate results when used with certain drugs.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device that may produce inaccurate diagnostic results due to drug interference, posing a risk of harm to patients through incorrect clinical decisions, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling the Dimension Glucose (GLUC) assay (Product Number: DF40/10444973). The recall involves 80,466 units distributed in all 50 U.S. states, the District of Columbia, Guam, Puerto Rico, and 108 foreign countries.
The recall is being conducted because various assays used to measure chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine. This interference can cause falsely elevated or falsely depressed results, which may impact clinical decision-making.
Healthcare providers and laboratories should stop using the affected Dimension Glucose assays and contact Siemens Healthcare Diagnostics for further instructions on the recall and replacement options.
The recalled product
- Product
- Dimension Glucose (GLUC) assay (Product Number: DF40/10444973) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 80,466
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All lots
Distribution
Part of a larger recall action
This product is one of 17 recalled by Siemens Healthcare Diagnostics, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 17 products in this recall →Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · inaccurate results
See all →- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- HighHealthmark Protein Test Recalled Due to Expired Ingredient
FDA (Devices) · 2026-09-02
- HighHardyDisk AST Cefiderocol 30µg for In Vitro Testing Recalled Due to Incorrect Cartridges
FDA (Devices) · 2026-08-12
- HighGEM Premier 7000 with iQM3 Blood Analysis Systems Recalled for Solution Concentration Issues
FDA (Devices) · 2026-08-12
Similar recalls
- HighSiemens Healthcare Recalls Dimension Assays: Triglycerides (TGL) Due to Interference
FDA (Devices) · 2016-05-18
- HighSiemens Uric Acid Assay Recalled Due to Drug Interference
FDA (Devices) · 2016-05-18
- SevereSiemens Uric Acid Assays Recalled Due to Drug Interference
FDA (Devices) · 2016-05-18
- HighSiemens Dimension Enzymatic Creatinine Assay Recalled for Interference
FDA (Devices) · 2016-05-18
- HighSiemens Dimension Vista Triglyceride Assay Recalled for Interference
FDA (Devices) · 2016-05-18
- HighSiemens ADVIA Chemistry Glucose Oxidase Assay Recalled for Interference
FDA (Devices) · 2016-05-18