Agency classClass IIOur severity:HighFDA (Devices)·Z-1610-2021·Announced 2021-05-26

Medtronic Guardian Sensor 3 Recall for Hydroxyurea Interaction

Medtronic is recalling Guardian Sensor 3 units worldwide due to a risk of inaccurate elevated glucose readings when used with hydroxyurea.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (inaccurate glucose monitoring) where injury has not yet been reported, fitting the High severity criteria.

Plain-English summary

Medtronic Inc. is recalling 54,211 units of the Guardian Sensor (3) (REF: MMT-7020C4) worldwide. The product is intended for use with Medtronic Diabetes glucose sensing systems to continuously monitor glucose levels in persons with diabetes. The recall applies to all lot numbers within the active shelf life.

The recall was initiated because using the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug, could result in inaccurate elevated sensor glucose readings. The source text notes that the product was not distributed within the US.

Consumers should stop using the affected sensors if they are actively taking hydroxyurea. Medtronic is the recalling firm for this Class II medical device recall.

The recalled product

Product
Medtronic REF: MMT-7020C4 Guardian Sensor (3) for the following packaging: GS3 5PK 2L CANADA/ GTIN: 20763000179612 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.…
Manufacturer
Medtronic Inc.
Affected units
54,211

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All Lot numbers within active shelf life

Distribution

Distribution scope not specified by the agency.

Part of a larger recall action

This product is one of 20 recalled by Medtronic Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 20 products in this recall →