Medtronic Guardian Real-Time Monitor Recalled for Finnish Translation Error
Medtronic MiniMed is recalling 254 Guardian Real-Time Monitors distributed in Finland due to a language translation error that reverses the on-screen labels for LOW and HIGH glucose alerts.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, which typically indicates a situation where use of the product may cause temporary or medically reversible adverse health consequences. The hazard is a labeling/translation error rather than a structural defect or contamination, fitting the 'Moderate' criteria for minor labeling errors or low-risk issues.
Plain-English summary
Medtronic MiniMed Inc. is recalling 254 units of the Guardian Real-Time Monitor, Model No. CSS7100. This medical device is indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin for adults (18 years and older) with diabetes mellitus.
The recall is due to a language translation error in the Finnish language version of the device. Specifically, the on-screen Finnish translation for the Predictive Alerts setting is reversed, displaying HIGH where LOW should appear and LOW where HIGH should appear. This error could impact the user's interpretation of glucose alerts.
The affected units were distributed in Finland. Consumers with this device should check their specific unit against the list of affected lot codes provided by the manufacturer or contact Medtronic MiniMed for further instructions on how to proceed.
The recalled product
- Product
- Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with diabetes mellitus, for the purpose of improving diabetes management.
- Manufacturer
- Medtronic MiniMed Inc.
- Hazard
- Affected units
- 254
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Medtronic MiniMed Inc.
See all →- HighMedtronic MiniMed 780G Insulin Pumps Recalled for Chemical Damage Risk
FDA (Devices) · 2026-08-26
- HighMedtronic MiniMed 640G Insulin Pumps Recalled for Chemical Damage Risk
FDA (Devices) · 2026-08-26
- SevereMedtronic MiniMed 670G Insulin Pumps Recalled Due to Chemical Damage Risk
FDA (Devices) · 2026-08-26
- HighMiniMed Go App Recalled Due to Missed Insulin Alerts
FDA (Devices) · 2026-07-29
- HighMiniMed 780G Insulin Pump Software Anomaly Recall
FDA (Devices) · 2026-07-15
Same hazard · mis labeling
See all →- ModerateOlympus Recalls PowerSeal Sealer Due to Incorrect Outer Box Labeling
FDA (Devices) · 2026-09-02
- ModerateAVID Medical Recalls Halyard Total Joint Kits for Latex Labeling Deficiency
FDA (Devices) · 2026-09-02
- HighByron White A-L Complex Herbal Supplement Recalled for Undeclared Walnut
FDA (Food) · 2026-09-02
- ModerateMedline Trach Tube Kit Recalled for Component Labeling Discrepancy
FDA (Devices) · 2026-09-02
- SevereWhite Vein Advanced Alkaloids Chewable Tablets Recall for Excess Active Ingredient
FDA (Food) · 2026-09-02