RayStation software recall due to virtual wedge field limitations
RaySearch Laboratories is recalling RayStation software because virtual wedge fields may not cover the entire field, resulting in a high dose plateau.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device used for radiation therapy where a software defect could result in unintended high dose delivery (high dose plateau), representing a risk of harm where injury has not yet been reported.
Plain-English summary
RaySearch Laboratories AB is recalling RayStation standalone software treatment planning systems, including model numbers 4.0 through 9.1. The recall involves 1,389 licenses distributed nationwide in the United States.
The recall is issued because, due to physical limitations of the treatment machine, the maximum field opening for Siemens Virtual Wedge fields is more restricted in the wedged direction than without a wedge. Delivering a larger field opening with a virtual wedge does not result in a wedged field shape over the entire field. Instead, a high dose plateau region occurs at the toe side of the field beyond the recommended field size.
Authorized intended users who use RayStation for radiation therapy and medical oncology should contact RaySearch Laboratories AB for further instructions on how to address this issue.
The recalled product
- Product
- RayStation standalone software treatment planning system, Model Numbers 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 7.0, 8.0, 8.1, 9.0, 9.1 - Product Usage: RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. Aft
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Hazard
- Affected units
- 1,389
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Systems with software versions: 4.0.0.14
- 4.0.1.4
- 4.0.2.9
- 4.0.3.4
- 4.5.0.19
- 4.5.1.14
- 4.5.2.7
- 4.7.0.15
- 4.7.1.10
- 4.7.2.5
- 4.7.3.13
- 4.7.4.4
- 4.7.5.4
- 4.7.6.7
- 4.9.0.42
- 5.0.0.37
- 5.0.1.11
- 5.0.2.35
- 5.0.3.17
- 6.0.0.24
Distribution
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · radiation dose error
See all →- HighSiemens Interventional Fluoroscopic X-Ray System Software Communication Failure
FDA (Devices) · 2025-07-09
- HighRayStation radiation therapy software: incorrect density uncertainty calculations in dose planning
FDA (Devices) · 2025-02-26
- HighRadiation Therapy Planning System May Report Dose Values Too High
FDA (Devices) · 2024-05-08
- HighPhilips Radiography System Firmware Defect Risks Incorrect Radiation Dose
FDA (Devices) · 2022-09-07
- HighRadiography System Firmware Flaw May Cause Incorrect Image Orientation and Dose
FDA (Devices) · 2022-07-20