Siemens SOMATOM Force CT Systems Recalled for Image Artifacts
Siemens is recalling 37 SOMATOM Force CT systems because a software error in the Neonate Head protocol may cause image artifacts, potentially leading to misdiagnosis.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a risk of misdiagnosis due to image artifacts, which is a significant risk of harm, but the source text does not report any specific injuries, hospitalizations, or fatalities.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 37 SOMATOM Force Computed Tomography (CT) x-ray systems. These systems are intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.
The recall is due to an incorrect parameterization of the reconstruction algorithm used for the pediatric head kernel (Hp38) in the Neonate Head protocol. This error can result in image artifacts that mimic non-existing blood or liquid or fail to depict actual existing blood or liquid. This may lead to misdiagnosis or errors in surgery planning.
The affected units were distributed nationwide in the United States, including in North Carolina, New Hampshire, Illinois, Massachusetts, Washington, Pennsylvania, Maryland, Minnesota, Texas, New York, South Carolina, Ohio, California, Kentucky, Montana, and Iowa. Healthcare facilities should contact Siemens for instructions on correcting the software parameterization.
The recalled product
- Product
- SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 37
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- SOMATOM Force Material # 10742326
- serial numbers 75550
- 75526
- 75512
- 75567
- 75536
- 75514
- 75563
- 75510
- 75454
- 75540
- 75571
- 75437
- 75500
- 75527
- 75475
- 75535
- 75481
- 75439
- 75450
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · misdiagnosis
See all →- HighPhilips EPIQ Elite Ultrasound System Recalled for Software Error
FDA (Devices) · 2026-09-30
- HighGE Healthcare Recalls Centricity PACS Software Due to Data Association Errors
FDA (Devices) · 2026-09-23
- HighGE Healthcare Recalls Centricity PACS Server V7 Due to Data Association Error
FDA (Devices) · 2026-09-23
- SevereNova Biomedical Recalls StatStrip Glucose Meters Due to Software Defect
FDA (Devices) · 2026-09-23
- HighGE Healthcare Enterprise Archive V8 Software Recalled for Data Association Errors
FDA (Devices) · 2026-09-23
Similar recalls
- HighSiemens CT System Recalled for Image Artifacts and Safety Issues
FDA (Devices) · 2015-03-18
- HighSiemens SOMATOM Force CT Scanner Recalled for Imaging Artifacts
FDA (Devices) · 2015-01-21
- HighSiemens SOMATOM Force CT Systems Recalled for Software Scan Abort Issues
FDA (Devices) · 2019-08-21
- ModerateSiemens Recalls SOMATOM Force CT Systems for Software Bugs
FDA (Devices) · 2016-12-14
- HighSiemens SOMATOM Force CT Systems Recalled for Software and Firmware Bugs
FDA (Devices) · 2015-10-14
- HighSiemens SOMATOM Drive CT Systems Recalled for Software Safety Issues
FDA (Devices) · 2019-08-21