Ohmeda Giraffe Incubator Recall Due to Power Cord Overheating Risk
Ohmeda Medical is recalling Giraffe Incubators because certain power cords may overheat, posing a fire or thermal injury risk.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as FDA Class II and involves a risk of fire and thermal injury, which aligns with the 'Severe' criteria for significant injury reports or high-risk hazards.
Plain-English summary
Ohmeda Medical is recalling Giraffe Incubators, which are infant incubators designed to provide a controlled temperature environment for neonates who cannot regulate their own body heat. The recall involves 5,993 units distributed worldwide, including 5,080 in the United States and Puerto Rico, and 913 in other countries such as the UK, Canada, and Brazil.
The recall is due to a defect in certain power cords manufactured by Electric-Cord, referred to as 'Taller Bridge' power cords. These cords may overheat where they connect to the wall outlet, which could result in fire, charring, smoke, or sparking. Such conditions could cause thermal injury to patients or staff.
Healthcare facilities should stop using the affected incubators and contact Ohmeda Medical for instructions on how to proceed. The source text does not specify a deadline or a specific method for consumers to report the issue, but immediate cessation of use is implied by the safety hazard.
The recalled product
- Product
- Giraffe Incubator The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed temperature controlled environment
- Manufacturer
- Ohmeda Medical
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- See Attached Spreadsheat
Distribution
Part of a larger recall action
This product is one of 11 recalled by Ohmeda Medical under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 11 products in this recall →Related recalls
Same manufacturer · Ohmeda Medical
See all →- HighGiraffe Omnibed Infant Incubator Recalled for Inadequate Cleaning Instructions
FDA (Devices) · 2019-10-16
- HighOhmeda Medical Recalls Rotating IV Poles Due to Fall Risk
FDA (Devices) · 2018-04-04
- SevereOhmeda Infant Warmer System Recalled Due to Overheating Power Cords
FDA (Devices) · 2017-03-29
- SevereOhmeda Infant Radiant Warmers Recalled for Power Cord Overheating Risk
FDA (Devices) · 2017-03-29
- SevereOhmeda Medical Recalls Giraffe Incubators Due to Power Cord Overheating
FDA (Devices) · 2017-03-29
Same hazard · fire
See all →- SevereVQQQ Gas Patio Heaters Recalled for Fire and Burn Hazards
CPSC · 2026-10-22
- HighAutocar Recalls Xpeditor Trucks for Fuel Hose Separation Risk
NHTSA · 2026-10-21
- SevereAudi Recalls 2024-2025 Q4 E-Tron Models Due to Battery Fire Risk
NHTSA · 2026-10-21
- HighForest River Recalls 2023-2026 Pause Travel Trailers for Wiring Defect
NHTSA · 2026-10-20
- SevereForest River Recalls 2026-2027 Cedar Creek Trailers for Fire Risk
NHTSA · 2026-10-16
Similar recalls
- HighGiraffe Omnibed Infant Incubator Recalled for Inadequate Cleaning Instructions
FDA (Devices) · 2019-10-16
- CriticalGE Healthcare Recalls Giraffe Infant Warmers Due to Latching Defect
FDA (Devices) · 2019-12-11
- SevereGE Healthcare Recalls Giraffe OmniBed and Incubator Units for Display Errors
FDA (Devices) · 2013-04-17
- SevereGE Healthcare Recalls Giraffe OmniBed Incubators Due to Oxygen Display Error
FDA (Devices) · 2020-06-24
- HighGE Healthcare Recalls Giraffe Infant Warmers Due to Loose Adjustment Knob
FDA (Devices) · 2014-02-12
- HighGE Healthcare Recalls Giraffe Incubators with Servo Oxygen Module
FDA (Devices) · 2020-06-24