The Recall Desk

Manufacturer

Ohmeda Medical

13 recalls in our database name Ohmeda Medical as the manufacturing or recalling firm.

What the numbers show

Total
13
Critical
0
Severe
11
Most recent
2019-10-16

Of the 13 recalls from Ohmeda Medical we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 11 (85%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–13 of 13

  • HighFDA (Devices)·Z-0041-2020·2019-10-16

    Giraffe Omnibed Infant Incubator Recalled for Inadequate Cleaning Instructions

    Ohmeda Medical is recalling Giraffe Omnibed infant incubators because cleaning guidelines failed to prevent debris buildup, creating a potential infection risk.

    Product
    Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same tim
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1181-2018·2018-04-04

    Ohmeda Medical Recalls Rotating IV Poles Due to Fall Risk

    Ohmeda Medical is recalling 1,179 rotating IV poles because they may fall, posing a risk of injury to patients, bystanders, or operators.

    Product
    Rotating IV Pole Rotating IV Pole falls under the category classified as capital equipment, an optional accessory that may be purchased to facilitate attachment and consolidation of patient equipment to one central patient care area. Devices commonly attached to IV poles inc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1510-2017·2017-03-29

    Ohmeda-Ohio CarePlus Incubator Recall Due to Power Cord Overheating Risk

    Ohmeda Medical is recalling CarePlus Incubators because certain power cords may overheat, posing a risk of fire or thermal injury.

    Product
    CarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1511-2017·2017-03-29

    Ohmeda Infant Warmer System Recalled Due to Overheating Power Cords

    Ohmeda Medical is recalling IWS Infant Warmer Systems because certain power cords may overheat, posing a fire and thermal injury risk.

    Product
    IWS Infant Warmer System-Infant radiant warmers provide infrared heat, in a controlled manner, to neonates who are unable to thermoregulate based on their own physiology or necessitate external heat to smoothen the transition from the uterus to the external environment. They ful
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1506-2017·2017-03-29

    Ohmeda Infant Radiant Warmers Recalled for Power Cord Overheating Risk

    Ohmeda Medical is recalling Giraffe and Panda infant radiant warmers because certain power cords may overheat, posing a fire or thermal injury risk.

    Product
    Giraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1505-2017·2017-03-29

    Ohmeda Medical Recalls Giraffe Incubators Due to Power Cord Overheating

    Ohmeda Medical is recalling Giraffe Incubators because certain power cords may overheat, posing a risk of fire or thermal injury.

    Product
    Giraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature con
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1501-2017·2017-03-29

    Ohmeda Giraffe Exam Light Recalled Due to Power Cord Overheating Risk

    Ohmeda Medical is recalling Giraffe Exam Lights because certain power cords may overheat, posing a risk of fire, smoke, or thermal injury.

    Product
    Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examination. It increases ambient light to facilitate performance of procedures on the mattress field. It is designed to mount on the accessory railings of the Giraff
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1503-2017·2017-03-29

    Ohmeda Medical Spot PT Lite Phototherapy System Recalled for Power Cord Fire Risk

    Ohmeda Medical is recalling Spot PT Lite Phototherapy Systems because certain power cords could overheat, posing a risk of fire or thermal injury.

    Product
    Giraffe Spot PT Ohmeda Medical Spot PT Lite Phototherapy System- The Spot PT Lite Phototherapy System provides light therapy for the treatment of hyperbilirubinemia, commonly known as neonatal jaundice, during the newborn period in the hospital
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1509-2017·2017-03-29

    Ohmeda Medical Recalls Giraffe OmniBed Infant Incubators Due to Fire Risk

    Ohmeda Medical is recalling Giraffe OmniBed infant incubators and warmers because certain power cords may overheat, posing a risk of fire or thermal injury.

    Product
    Giraffe OmniBed Ohmeda Medical Omnibed- The OmniBed is a combination of an infant incubator ( incubator ) and an infant radiant warmer ( warmer ). Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiolo
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1508-2017·2017-03-29

    Ohmeda Infant Radiant Warmers Recalled for Power Cord Overheating Risk

    Ohmeda Medical is recalling Giraffe and Panda infant radiant warmers because certain power cords could overheat, posing a risk of fire or thermal injury.

    Product
    Giraffe Bedded Warmer, Panda Warmer Modifiction to Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neon
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1504-2017·2017-03-29

    Ohmeda Medical Recalls Giraffe OmniBed Infant Incubators Due to Fire Risk

    Ohmeda Medical is recalling Giraffe OmniBed infant incubators and warmers because certain power cords may overheat, posing a risk of fire or thermal injury.

    Product
    Giraffe OmniBed Ohmeda Medical Giraffe OmniBed- The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1507-2017·2017-03-29

    Ohmeda Infant Warmers Recalled Due to Power Cord Overheating Risk

    Ohmeda Medical is recalling Giraffe and Panda infant radiant warmers because certain power cords may overheat, posing a fire or thermal injury risk.

    Product
    Giraffe Bedded Warmer, Panda Warmer Modification to Giraffe and Panda Warmers (Giraffe and Panda Uninterruptible Power Supply) - Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infan
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1502-2017·2017-03-29

    Ohmeda Giraffe Incubator Recall Due to Power Cord Overheating Risk

    Ohmeda Medical is recalling Giraffe Incubators because certain power cords may overheat, posing a fire or thermal injury risk.

    Product
    Giraffe Incubator The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed temperature controlled environment
    Category
    Medical Device
    Distribution
    Distributed nationwide