GE Healthcare Recalls Giraffe OmniBed and Incubator Units for Display Errors
GE Healthcare is recalling 679 Giraffe OmniBed and Incubator units because display boards may randomly overwrite stored information and set points upon power up or recovery from a power failure.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as FDA Class I, which the rubric defines as a situation where use of the product may cause serious adverse health consequences or death. The defect involves the overwriting of critical oxygen and heater set points for neonatal patients.
Plain-English summary
GE Healthcare, LLC is recalling 679 units of the Giraffe OmniBed and Giraffe Incubator (p/n 6600-0723-700, rev A and B). The affected devices are combination infant incubators and warmers. The recall covers units distributed worldwide, including in the United States and numerous international countries.
The recall is being issued because the neonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure. This defect could result in the device not maintaining the intended settings for the infant.
Healthcare facilities should check their inventory for the specific serial numbers listed in the recall notice. If a unit is affected, it should be removed from service and GE Healthcare should be contacted for instructions on repair or replacement.
The recalled product
- Product
- Giraffe OmniBed¿ and Giraffe Incubator p/n 6600-0723-700 (rev A and B) labeled in part: ***GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com***. GE Healthcare Giraffe Omnibed is a combination of an infant incubator and an infant warmer.
- Manufacturer
- GE Healthcare, LLC
- Category
- Medical Device — Neonatal Care
- Affected units
- 679
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- OmniBed Serial Numbers: HDGR52546
- HDGR52736
- HDGR52802
- HDGR52861
- HDGR52910
- HDGR52958
- HDGR53007
- HDGR53058
- HDGR53110
- HDGR53185
- HDGR52547
- HDGR52737
- HDGR52803
- HDGR52862
- HDGR52911
- HDGR52959
- HDGR53008
- HDGR53059
- HDGR53112
- HDGR53186
Distribution
Related recalls
Same manufacturer · GE Healthcare, LLC
See all →- HighGE SIGNA Premier MRI magnets may have ferrous fittings fitted
FDA (Devices) · 2026-04-08
- HighGE Medical Imaging Software: Inaccurate Measurements in Global Stack Viewport
FDA (Devices) · 2022-08-31
- HighGE Healthcare CT, PET, and NM Patient Tables May Drop During Service
FDA (Devices) · 2022-08-17
- HighGE Centricity PACS Software fails to notify of exam study modifications
FDA (Devices) · 2022-08-03
- SevereGE Healthcare Anesthesia Systems Recalled for Potential Transducer Connection Reversal
FDA (Devices) · 2022-07-27
Same hazard · device malfunction
See all →- SevereMedtronic MiniMed Flex Insulin Pumps Recalled for Failed Verification Tests
FDA (Devices) · 2026-10-14
- ModerateWilson-Cook Recalls Hemospray Endoscopic Hemostat Due to Adherence Issues
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Et Control Software Recalled for Anesthetic Delivery Interruption
FDA (Devices) · 2026-09-23
Similar recalls
- SevereGE Healthcare Recalls Giraffe OmniBed Incubators Due to Oxygen Display Error
FDA (Devices) · 2020-06-24
- CriticalGE Healthcare Recalls Giraffe Infant Warmers Due to Latching Defect
FDA (Devices) · 2019-12-11
- HighGE Healthcare Recalls Giraffe OmniBed Neonatal Incubators Due to Electric Leakage
FDA (Devices) · 2018-11-28
- HighGiraffe OmniBed neonatal incubator recall due to incorrect secondary latch installation
FDA (Devices) · 2023-07-12
- SevereGE Healthcare Recalls Giraffe OmniBed Infant Warmers Due to Latch Defect
FDA (Devices) · 2019-12-11
- SevereGE Healthcare Recalls Giraffe OmniBed Carestation CS1 Due to Fastener Risk
FDA (Devices) · 2019-11-13