The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1493-2015·Announced 2015-04-29

Wright Medical Recalls Crosscheck Plating System Due to Specification Defects

Wright Medical Technology is recalling 172 units of the Crosscheck Plating System because specific dimensions are out of specification, posing a risk of intraoperative malfunction.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (intraoperative malfunction) where injury has not yet been reported, fitting the criteria for a High severity score.

Plain-English summary

Wright Medical Technology, Inc. is recalling 172 units of the CROSSCHECK(R) Plating system, MTP CX Plate, Right (REF CCP-MPX1R). This orthopedic device was distributed nationwide. The affected lot numbers are 35049-15A, 35049-15A-1, 35864, 358641, 362772, and 36277-2-1.

The recall was initiated because the minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification. These defects may result in intraoperative malfunction during surgical procedures.

Healthcare providers and patients who have received or used these specific devices should contact Wright Medical Technology for further instructions on how to proceed.

The recalled product

Product
CROSSCHECK(R) Plating system, MTP CX Plate, Right, REF CCP-MPX1R. Orthopedic use.
Affected units
172

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (6)

  • Lot Numbers: 35049-15A
  • 35049-15A-1
  • 35864
  • 358641
  • 362772
  • 36277-2-1

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 19 recalled by Wright Medical Technology, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 19 products in this recall →