The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1371-2019·Announced 2019-05-29

Polymer Technology Systems Recalls Lipid Panel Test Strips Due to Fit Issues

Polymer Technology Systems is recalling Lipid Panel test strips because they may not fit tightly into the optical block, potentially causing incorrect results.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II and involves a risk of harm (incorrect medical results) where injury has not yet been reported, fitting the High severity criteria.

Plain-English summary

Polymer Technology Systems, Inc. is recalling Lipid Panel test strips, REF 1710, which are also private labeled under Henry Schein as REF 570-0414. The recall involves 67,643 units that were distributed domestically nationwide.

The recall was initiated because some lots of the test strips do not fit tightly into the optical block. This defect may cause the strips to fall out during analysis, which could lead to an incorrect result being displayed.

Consumers and healthcare providers should stop using the affected test strips and contact Polymer Technology Systems for further instructions.

The recalled product

Product
Lipid Panel test strips, REF 1710, also private labeled under Henry Schein as REF 570-0414
Affected units
67,643

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • P832

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 3 recalled by Polymer Technology Systems, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 3 products in this recall →

Similar recalls