Neurotherm NT2000iX RF Generator Recalled for Uncleared Software Settings
Neurotherm is recalling 25 NT2000iX RF Generators because they shipped with international software settings enabled, allowing access to modes not cleared for use in the United States.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any hospitalizations, injuries, or illnesses. The hazard is related to the availability of uncleared software modes rather than a direct physical defect causing immediate harm, fitting the criteria for a moderate severity score.
Plain-English summary
Neurotherm, Inc. is recalling 25 units of the Neurotherm NT2000iX RF Generator (Model Number: RFG-NT-2000iX). The recall was initiated because the devices were shipped with the international setting turned on. This configuration allows access to Corodotomy, Bi-Polar, and No Temperature modes, which are not cleared by the FDA for use in the United States.
The affected units were distributed in California, Colorado, Iowa, Maine, New Jersey, New York, South Carolina, Texas, and Wisconsin. Specific serial numbers for the recalled units are listed in the official recall record. The agency has classified this as a Class II recall.
Healthcare providers and consumers who possess these devices should stop using them and contact Neurotherm, Inc. for instructions on how to resolve the issue. The recall is necessary to prevent the use of software modes that have not received FDA clearance for the U.S. market.
The recalled product
- Product
- Neurotherm NT2000iX RF Generator Model Number: RFG-NT -2000iX
- Manufacturer
- Neurotherm, Inc.
- Affected units
- 25
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial Numbers: 2593-13
- 2717-13
- 2616-13
- 2704-13
- 2735-13
- 2360-12
- 2732-13
- 2727-13
- 2284-11
- 2726-13
- 2728-13
- 2426-12
- 2700-13
- 2508-12
- 2643-13
- 2733-13
- 2730-13
- 2724-13
- 2453-12
- 2631-13
Distribution
Related recalls
Same manufacturer · Neurotherm, Inc.
See all →- ModerateNeurotherm Curved Radiofrequency Cannula Recalled for Mislabeling
FDA (Devices) · 2016-03-30
Same hazard · software error
See all →- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- HighPhilips Spectral CT 7500 Software Recall for Radiation Exposure Risk
FDA (Devices) · 2026-10-07
- HighBard EnCor EnCompass Breast Biopsy System Software Recall
FDA (Devices) · 2026-10-07
- ModerateBeckman Coulter AU480 Chemistry Analyzer Software Error Recall
FDA (Devices) · 2026-10-07
- HighPhilips EPIQ Elite Ultrasound System Recalled for Software Error
FDA (Devices) · 2026-09-30
Similar recalls
- HighAbbott NeuroTherm Radiofrequency Generator Recalled for Capacitor Failure
FDA (Devices) · 2021-05-19
- ModerateFDA Class II Quantum 2000 Electrosurgical Generators Voluntary Recall
FDA (Devices) · 2021-11-24
- ModerateGentherm Medical Recalls Blanketrol II Thermal Regulating Systems
FDA (Devices) · 2026-08-26
- ModerateAbbott NT2000iX radiofrequency generator serviced with out-of-tolerance tool
FDA (Devices) · 2023-11-08
- ModerateFujifilm Recalls 14 MXR-35 X-ray Generators for Missing Certification Labeling
FDA (Devices) · 2014-05-28
- HighBiosense Webster Recalls SMARTABLATE RF System Generator Units
FDA (Devices) · 2017-12-13