Abbott NT2000iX radiofrequency generator serviced with out-of-tolerance tool
Abbott is recalling 34 NT2000iX radiofrequency generators that were serviced using an out-of-tolerance calibration tool, which may have provided inaccurate measurement results for capacitance failures.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: This is an FDA Class II recall with no reported illnesses or injuries. The hazard is theoretical—potential inaccuracy in measurement results from an out-of-tolerance servicing tool. No direct patient harm has been reported, making this a moderate-level precautionary recall.
Plain-English summary
Abbott is recalling 34 NT2000iX Radiofrequency (RF) generators, model RFG-NT-2000, used for lesioning procedures. These devices were serviced using a calibration tool that was subsequently found to be out of tolerance. As a result, the servicing may not have provided accurate measurement results for potential capacitance failures.
The affected devices were distributed in the United States, Taiwan, and Colombia. Affected units are identified by the UDI/DI 05415067022417 and associated serial numbers available from Abbott.
Clinicians and facility managers with affected equipment should contact Abbott for additional information and guidance regarding their devices.
The recalled product
- Product
- Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator
- Manufacturer
- Abbott
- Hazard
- calibration-error
- measurement-inaccuracy
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- UDI/DI 05415067022417
- serial numbers: 16930509
- 16983330
- 17583407
- 17634837
- 17783928
- 17783930
- 17905961
- 17928413
- 17954521
- 17991999
- 18047384
- 18493711
- 18522937
- 18675068
- 18690920
- 18709734
- 18742997
- 2475-12
- 2504-12
Distribution
Distribution scope not specified by the agency.
Related recalls
Same category
- HighMedtronic DLP One-Piece Pediatric Arterial Cannula Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- SevereMedtronic Bio-Medicus Adult Cannula Kit Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- SevereT-Lok Bone Marrow Tray Kits Recalled for Sterility Testing Failure
FDA (Devices) · 2026-07-22
- SevereAbiomed Impella CP Smart Assist Set Heart Pump Introducer Sheath Leakage
FDA (Devices) · 2026-07-22