Surgeon console displays recalled over firmware that can lose visualization
Intuitive Surgical is recalling 47 da Vinci 5 Surgeon Console Viewer Displays manufactured with outdated firmware, which may cause loss of the 3D surgical field or complete visualization loss.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: the firm states complete visualization loss with instruments still active could lead to instrument contact with tissue causing injury or bleeding, and no injuries are reported.
Plain-English summary
The da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, part number 380730-45, is being recalled by Intuitive Surgical, Inc.
The record states the robotic-assisted surgical system has a Viewer Display manufactured with outdated firmware. It states this may result in loss of one display, meaning loss of the 3D surgical field and depth perception with 2D maintained, or loss of both displays, meaning complete visualization loss, without fault. The record states instruments remain active, which could lead to instrument contact with tissue resulting in injury or bleeding, and that the situation may require an alternate surgical modality.
The record lists 47 units under UDI 00886874119747, with serial numbers recorded in the enforcement record. Distribution was nationwide in the United States across the states listed in the record. The FDA has classified the action as Class II.
The enforcement record reports no injuries. Facilities should follow Intuitive Surgical's instructions.
The recalled product
- Product
- da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45
- Manufacturer
- Intuitive Surgical, Inc.
- Affected units
- 47
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- UDI: 00886874119747
- Serial: SQ0326
- SQ0336
- SQ0341
- SQ0345
- SQ0346
- SQ0356
- SQ0367
- SQ0368
- SQ0375
- SQ0377
- SQ0379
- SQ0380
- SQ0386
- SQ0387
- SQ0389
- SQ0394
- SQ0395
- SQ0398
- SQ0401
Distribution
Related recalls
Same manufacturer · Intuitive Surgical, Inc.
See all →- Moderateda Vinci SP Instrument Arm Drape may detach during surgical procedures
FDA (Devices) · 2026-07-22
- SevereDa Vinci surgical system operative components with E-brake retainer pin defect
FDA (Devices) · 2026-07-01
- HighDa Vinci 5 tower may unintentionally activate the infrared handheld camera
FDA (Devices) · 2026-06-10
- Highda Vinci X and Xi Surgical Systems Recalled for Manipulator Arm Screw Breakage
FDA (Devices) · 2026-05-13
- SevereIntuitive 8 mm SureForm 30 Gray Reloads Incomplete Staple Defect
FDA (Devices) · 2026-05-13
Similar recalls
- HighDa Vinci 5 Surgical System Console Recalled Due to Software Error
FDA (Devices) · 2026-02-25
- HighSurgical robot console motors recalled over an error that locks ergonomics
FDA (Devices) · 2025-12-31
- HighIntuitive Surgical Recalls da Vinci Si Surgical Systems Due to Mounting Risk
FDA (Devices) · 2013-11-06
- LowIntuitive Surgical da Vinci Si Surgeon Console Labeling Clarification
FDA (Devices) · 2013-06-05
- HighIntuitive Surgical Recalls da Vinci Xi Systems Due to Defective Capacitors
FDA (Devices) · 2020-02-12
- HighDa Vinci 5 surgical console foot pedals recalled for spring failure
FDA (Devices) · 2025-03-19