Philips DigiTrak XT Holter Recorder Software Recall for Malfunction
Philips is recalling DigiTrak XT Holter Recorders because a software defect causes the device to fail when an AAA battery is present.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard is a functional failure (Error 602) that prevents the device from operating for new studies, but the source text does not report any injuries, illnesses, or deaths.
Plain-English summary
Philips North America, LLC is recalling 44,442 units of the DigiTrak XT Holter Recorder, Model Number 860322. The recall affects software versions B.1 and C.2. The devices are distributed in the United States and numerous foreign countries.
The recall is due to a software defect that causes the recorder to display "Error: 602" and fail to function for a new patient study if an AAA battery is inserted and the user attempts to start the device, or if the recorder is inserted in the docking station. This malfunction prevents the device from performing its intended function for new patient studies.
All Philips 860322 DigiTrak XT Holter Recorders manufactured or distributed before December 31, 2019, are affected, regardless of serial number, ship date, or date of manufacture. Users should contact Philips North America, LLC for instructions on how to resolve the issue.
The recalled product
- Product
- Philips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available in four different configurations. Service tag numbers for these configurations are as follows: - 453564067151 24-Hour Configuration - 453564067161 48-Hour Config
- Manufacturer
- Philips North America, LLC
- Affected units
- 44,442
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (7)
- All Philips 860322 DigiTrak XT Holter Recorder
- manufactured or distributed before December 31
- 2019
- are affected
- regardless of serial number
- ship date
- and date of manufacture.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Philips North America, LLC
See all →- HighPhilips Recalls Cardiac Workstation 7000 Due to Inaccurate Heart Rate Data
FDA (Devices) · 2026-08-26
- HighPhilips Recalls Cardiac Workstation 5000 Units for Inaccurate Heart Rate Data
FDA (Devices) · 2026-08-26
- ModeratePhilips Telemetry Monitor 5500 Device Reset Risk Recall
FDA (Devices) · 2026-07-01
- ModeratePhilips Avalon Fetal Monitor FM20 Speaker Connector Assembly Recall
FDA (Devices) · 2026-07-01
- HighPhilips Avalon Fetal Monitor FM 30 Speaker Connector Assembly
FDA (Devices) · 2026-07-01
Same hazard · software defect
See all →- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25