Philips DigiTrak XT Holter Recorder Software Recall for Malfunction
Philips is recalling DigiTrak XT Holter Recorders because a software defect causes the device to fail when an AAA battery is present.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard is a functional failure (Error 602) that prevents the device from operating for new studies, but the source text does not report any injuries, illnesses, or deaths.
Plain-English summary
Philips North America, LLC is recalling 44,442 units of the DigiTrak XT Holter Recorder, Model Number 860322. The recall affects software versions B.1 and C.2. The devices are distributed in the United States and numerous foreign countries.
The recall is due to a software defect that causes the recorder to display "Error: 602" and fail to function for a new patient study if an AAA battery is inserted and the user attempts to start the device, or if the recorder is inserted in the docking station. This malfunction prevents the device from performing its intended function for new patient studies.
All Philips 860322 DigiTrak XT Holter Recorders manufactured or distributed before December 31, 2019, are affected, regardless of serial number, ship date, or date of manufacture. Users should contact Philips North America, LLC for instructions on how to resolve the issue.
The recalled product
- Product
- Philips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available in four different configurations. Service tag numbers for these configurations are as follows: - 453564067151 24-Hour Configuration - 453564067161 48-Hour…
- Manufacturer
- Philips North America, LLC
- Affected units
- 44,442
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (7)
- All Philips 860322 DigiTrak XT Holter Recorder
- manufactured or distributed before December 31
- 2019
- are affected
- regardless of serial number
- ship date
- and date of manufacture.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Philips North America, LLC
See all →- HighPhilips Spectral CT 7500 Software Recall for Radiation Exposure Risk
FDA (Devices) · 2026-10-07
- ModeratePhilips Achieva 1.5T MRI Systems Recalled for Software Memory Issue
FDA (Devices) · 2026-09-30
- ModeratePhilips MR 7700 MRI Systems Recalled for Software Issue
FDA (Devices) · 2026-09-30
- ModeratePhilips Ingenia MRI Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
- ModeratePhilips MRI Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
Same hazard · software defect
See all →- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- HighPhilips Spectral CT Systems Recalled for Software Issues Affecting Safety
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- ModerateUnited Imaging uCT 760 CT System Recalled for Software Visibility Issue
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uCT 550 CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
Similar recalls
- HighPhilips HeartStart XL+ Defibrillator Recall for Software Communication Failure
FDA (Devices) · 2014-04-16
- ModeratePhilips DigitalDiagnost Software Recall for X-ray System Stitching Defects
FDA (Devices) · 2018-10-31
- ModerateBraemar DL900 Holter Monitor Recalled for Operational Failure
FDA (Devices) · 2020-03-18
- SeverePhilips HeartStart XL+ Defibrillator Recall Due to Switch Failure
FDA (Devices) · 2019-11-27
- ModerateBraemar DL950 Holter Monitor Recall Due to Startup Error
FDA (Devices) · 2020-03-18
- ModeratePhilips DigitalDiagnost X-Ray System Recalled for Missing Exposure Beep Signal
FDA (Devices) · 2015-05-27