CircuitGuard Bacterial/Viral Filter Recalled for Plastic Connector Obstruction
Kung Shin Plastics is recalling 30,000 units of CircuitGuard Bacterial/Viral Filter with Elbow (Model SK200FPE/7056) due to plastic occlusion in the connector that prevents airflow and causes ventilation bag malfunction, risking oxygen deprivation and respiratory distress.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall of a medical device with serious risk of harm through oxygen deprivation and respiratory distress if ventilation fails. No illnesses or injuries are reported in the source text, making this a risk-of-harm situation without confirmed injury, which corresponds to Score 3 (High) per the rubric.
Plain-English summary
Kung Shin Plastics Co. Ltd. is recalling the CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalog number SK200FPE / 7056). Approximately 30,000 units were distributed in the United States in Illinois across 300 cases. The affected lot is C2N24002.
The machine-side connector of the filter may be occluded (blocked) by plastic, which prevents proper airflow. This defect can cause the ventilation bag to malfunction, making it difficult or impossible for users to properly inflate or deflate the bag during manual ventilation.
Users of this filter face potential health risks if the connector obstruction prevents proper ventilation. These risks include oxygen deprivation, respiratory distress, and associated medical complications. The recall was issued by the U.S. Food and Drug Administration as a Class II recall.
The recalled product
- Product
- CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)
- Manufacturer
- Kung Shin Plastics Co. Ltd.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI (01)00810050853682(10)C2N24002(11)240828(17)290827 Lot C2N24002
Distribution
Part of a larger recall action
This product is one of 2 recalled by Kung Shin Plastics Co. Ltd. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Kung Shin Plastics Co. Ltd.
See all →- HighHeat Moisture Exchanger Connector Occlusion May Impair Ventilation
FDA (Devices) · 2025-03-12
Other medical device recalls
- ModerateMedline Medical Convenience Kits Recalled for Adhesive Removal Issues
FDA (Devices) · 2026-10-28
- ModerateBoston Scientific Recalls Direxion Torqueable Microcatheters for Coating Defects
FDA (Devices) · 2026-10-28
- ModerateBoston Scientific Recalls Direxion Torqueable Microcatheters Due to Coating Defects
FDA (Devices) · 2026-10-28
- ModerateMedtronic NDI Passive Spheres Recalled for Reflective Material Variation
FDA (Devices) · 2026-10-28
- ModerateBoston Scientific Recalls Defective Direxion Torqueable Microcatheters
FDA (Devices) · 2026-10-28
Similar recalls
- SevereIntersurgical Flexible Breathing Circuit Recall Due to Exhalation Port Blockage Risk
FDA (Devices) · 2018-07-04
- HighPall Medical Breathing Circuit Filter Recall for Possible Air Leak
FDA (Devices) · 2014-07-16
- HighMechanical respiratory filter recalls due to sterilization process deviation
FDA (Devices) · 2021-06-23
- HighOxygenator devices recalled due to packaging damage compromising sterile barrier
FDA (Devices) · 2023-06-07
- SevereCardinal Health Recalls Presource Gyn Laparoscopy Kits Due to Airflow Obstruction Risk
FDA (Devices) · 2013-05-15
- SevereCardinal Health Recalls Presource Total Joint Kits Due to Airflow Obstruction Risk
FDA (Devices) · 2013-05-15