The Recall Desk

FDA (Devices) recall action 96224

Kung Shin Plastics Co. Ltd. recalled 2 products on 2025-03-12

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-03-12
Critical
0
Units

Why these products were recalled

The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2025-03-12

    Heat Moisture Exchanger Connector Occlusion May Impair Ventilation

    Kung Shin Plastics' CircuitGuard ThermoFlo Filter may have plastic obstructing the machine-side connector, preventing proper ventilation bag operation and risking oxygen deprivation during manual ventilation.

    Product
    CircuitGuard ThermoFlo Filter -Combined Heat and Moisture Exchanger, Bacterial Viral Filter w O2 Port and Elbow (Model/Catalogue number: SK200CPE / 6126)
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2025-03-12

    CircuitGuard Bacterial/Viral Filter Recalled for Plastic Connector Obstruction

    Kung Shin Plastics is recalling 30,000 units of CircuitGuard Bacterial/Viral Filter with Elbow (Model SK200FPE/7056) due to plastic occlusion in the connector that prevents airflow and causes ventilation bag malfunction, risking oxygen deprivation and respiratory distress.

    Product
    CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)
    Category
    Distribution
    1 state