The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1258-2018·Announced 2018-04-04

Encore Medical Recalls Altivate Anatomic Pegged Glenoid Trial Instruments

Encore Medical is recalling 92 units of Altivate Anatomic Pegged Glenoid Trials because the peg base design is larger on the trials than on the implants.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II and involves a design defect in a surgical implant trial that could lead to improper fit or surgical complications, fitting the criteria for a risk-of-harm product where injury has not yet been reported.

Plain-English summary

Encore Medical, Lp is recalling 92 units of the Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 46mm, REF 804-07-460. The recall covers all serial numbers and involves US Nationwide Distribution.

The recall was initiated after it was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial is much larger on the trials than on the implants. The affected part numbers include 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540. The AltiVate Anatomic Shoulder System is indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.

Healthcare providers should stop using the affected instruments and contact Encore Medical for further instructions.

The recalled product

Product
Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 46mm, REF 804-07-460 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
Manufacturer
Encore Medical, Lp
Affected units
92

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All Serial Numbers

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 5 recalled by Encore Medical, Lp under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 5 products in this recall →

Similar recalls