The Recall Desk

FDA (Devices) recall action 79241

Encore Medical, Lp recalled 5 products on 2018-04-04

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2018-04-04
Critical
0
Units
461

Why these products were recalled

It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–5 of 5

  • SevereFDA (Devices)··2018-04-04

    Encore Medical Recalls Altivate Anatomic Pegged Glenoid Trial Instruments

    Encore Medical is recalling 93 units of Altivate Anatomic Pegged Glenoid Trials because the peg base design is larger on the trials than on the implants.

    Product
    Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 54mm, REF 804-07-540 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Encore Medical Recalls Altivate Anatomic Pegged Glenoid Trial Instruments

    Encore Medical is recalling 92 units of Altivate Anatomic Pegged Glenoid Trials because the peg base design is larger on the trials than on the implants.

    Product
    Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Encore Medical Recalls Altivate Anatomic Pegged Glenoid Trial Instruments

    Encore Medical is recalling 92 units of Altivate Anatomic Pegged Glenoid Trials because the peg base design is larger on the trials than on the implants.

    Product
    Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 38mm, REF 804-07-380 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Encore Medical Recalls Altivate Anatomic Pegged Glenoid Trial Instruments

    Encore Medical is recalling 92 units of Altivate Anatomic Pegged Glenoid Trials because the peg base design is larger on the trials than on the implants.

    Product
    Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 46mm, REF 804-07-460 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Encore Medical Recalls Altivate Anatomic Pegged Glenoid Trial Instruments

    Encore Medical is recalling 92 units of Altivate Anatomic Pegged Glenoid Trials because the peg base design is larger on the trials than on the implants.

    Product
    Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 42mm, REF 804-07-420 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
    Category
    Distribution
    0 states