Synthes Radial Head Prosthesis Recalled Due to Potential Component Separation
Synthes USA is recalling 417 trial radial head prostheses because the trial head may come loose from the implant stem during surgery. No adverse events have been reported.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (medical device) where the hazard is theoretical and the source explicitly states that no injuries or adverse events have been reported, which aligns with the High severity criteria.
Plain-English summary
Synthes USA HQ, Inc. is recalling 417 units of the Synthes Radial Head Prosthesis: Trial Radial Head. This medical device is a two-piece modular system used for elbow joint replacement, consisting of a titanium alloy stem and cobalt chrome components. The recall covers all lots of specific part numbers listed in the agency documentation.
The recall was initiated because the trial head may come loose from the implant stem during manipulation of the arm during surgery. The source text explicitly states that no adverse events have been reported related to this issue and that the recall is not being initiated as a result of adverse events.
The affected devices were distributed nationwide in the United States, including in Illinois, Indiana, Massachusetts, Michigan, Missouri, New Jersey, New York, Ohio, Oregon, Pennsylvania, South Carolina, Texas, Virginia, Washington, and Wisconsin. Healthcare providers and patients should contact Synthes USA HQ, Inc. for further instructions regarding the return or replacement of these specific trial components.
The recalled product
- Product
- Synthes Radial Head Prosthesis: Trial Radial Head Elbow joint prosthesis: Synthes Radial Head Prosthesis System is a two-piece modular system comprised of titanium alloy stem and cobalt chrome components with an integral screw and side-loading application to allow in-situ…
- Manufacturer
- Synthes USA HQ, Inc.
- Affected units
- 417
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- All Lots of Part Numbers: 03.402.018
- 03.402.020
- 03.402.022
- 03.402.024
- 03.402.026
- 03.402.028
- 03.402.218
- 03.402.220
- 03.402.222
- 03.402.224
- 03.402.226
- 03.402.228
- 03.402.418
- 03.402.420
- 03.402.422
- 03.402.424
- 03.402.426
- 03.402.428
- 03.402.618
- 03.402.620
Distribution
Related recalls
Same manufacturer · Synthes USA HQ, Inc.
See all →- ModerateSynthes Electric Pen Drive Hand Switch Recalled for Labeling Error
FDA (Devices) · 2014-01-08
- ModerateSynthes Trochanteric Fixation Nail Recalled for Incorrect Packaging Labeling
FDA (Devices) · 2013-12-18
- ModerateSynthes TomoFix Osteotomy System Recalled for Surgical Technique Guide Labeling Error
FDA (Devices) · 2013-12-18
- SevereSynthes Recalls Titanium Trochanteric Fixation Nail Systems Due to Missing Bend
FDA (Devices) · 2013-12-11
- ModerateSynthes Soft Tissue Retractor Recalled for Locking Nut Malfunction
FDA (Devices) · 2013-12-11
Same hazard · component separation
See all →- HighInternational truck wiper linkage can separate causing wipers to fail
NHTSA · 2025-04-24
- SevereCadillac LYRIQ stabilizer bar bracket bolts may loosen and separate
NHTSA · 2025-04-10
- SevereAcura ZDX stabilizer bar bracket bolts may loosen and separate
NHTSA · 2025-04-10
- HighStryker Spine Monterey AL Implant Inserter Recall Due to Potential Gold Button Separation
FDA (Devices) · 2025-04-09
- HighStryker Spine Monterey AL Implant Inserter Recalled for Potential Unlock Button Separation
FDA (Devices) · 2025-04-09
Similar recalls
- HighSynthes Radial Head Prosthesis Recalled for Potential Stem Loosening
FDA (Devices) · 2017-02-08
- HighSkeletal Dynamics Recalls Align Radial Stems Due to Fracture Reports
FDA (Devices) · 2014-12-17
- SevereSynthes Recalls Titanium Polyaxial Reduction Heads for Spine Screws
FDA (Devices) · 2014-12-03
- HighSynthes Polyaxial Screw Body Recalled for Potential Post-Operative Loosening
FDA (Devices) · 2013-12-04
- HighSynthes Matrix Coronal Bender Recall Due to Tip Breakage
FDA (Devices) · 2013-12-04
- SevereSynthes Sternal Zip Flex System Recalled for Implant Cutting Risk
FDA (Devices) · 2013-11-06