The Recall Desk
ModerateFDA (Devices)·Z-1225-2013·Announced 2013-05-15

Stryker Irrigator Reusable Tips Recalled for Incorrect Sterilization Instructions

Stryker Endoscopy is recalling 147,921 reusable irrigator tips because the instructions for use listed incorrect pre-vacuum steam sterilization temperatures.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, but the source text identifies the issue as a labeling error regarding sterilization temperatures without reporting any specific illnesses, injuries, or hospitalizations, fitting the criteria for a moderate severity voluntary precautionary recall.

Plain-English summary

Stryker Endoscopy is recalling 147,921 units of Stryker Irrigator Reusable Tips, specifically the 5mm X 32cm Aspiration Needle Tip (Manufacturer Part Number 250-070-437). These suction irrigator tips are used to aid in cleaning and clearing the operative site during laparoscopic surgery.

The recall was initiated because the Instructions for Use incorrectly listed the pre-vacuum steam sterilization (wrapped method) temperature as 121-123°C, rather than the correct 132-133°C. This labeling error could potentially affect the proper sterilization of the device if followed by users.

The affected products were distributed worldwide, including in the United States and numerous international markets such as Canada, Mexico, and countries in Europe, Asia, and South America. Healthcare facilities and users should stop using the affected tips and contact Stryker Endoscopy for further instructions.

The recalled product

Product
Stryker Irrigator Reusable Tips: 5mm X 32cm Aspiration Needle Tip. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
Manufacturer
Stryker Endoscopy
Affected units
147,921

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Manufacturer Part Number: 250-070-437

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 20 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 20 products in this recall →