Stryker Laparoscopic Needle Holder Recall Due to Sterilization Instructions
Stryker Endoscopy is recalling 180,573 laparoscopic needle holders because the instructions for use do not support specific sterilization methods.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II and involves a documentation issue regarding sterilization instructions, which aligns with the rubric for moderate severity involving minor labeling or documentation errors.
Plain-English summary
Stryker Endoscopy is recalling 180,573 units of the PKG, 5MM SHAFT, NEEDLE HOLDER CURVED LEFT 45CM (P/N 0250080589). These laparoscopic manual instruments are intended for cutting, holding, grasping, and manipulating tissue and surgical auxiliaries. The recall is classified as Class II by the FDA.
The recall was initiated because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods. This documentation issue may affect the proper sterilization of the devices.
The affected units were distributed worldwide, including nationwide in the United States and internationally to numerous countries. Healthcare facilities and users should stop using the affected instruments and contact Stryker Endoscopy for further instructions regarding the recall.
The recalled product
- Product
- PKG, 5MM SHAFT, NEEDLE HOLDER CURVED LEFT 45CM, P/N 0250080589 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
- Manufacturer
- Stryker Endoscopy
- Affected units
- 180,573
Distribution
Part of a larger recall action
This product is one of 207 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 207 products in this recall →Related recalls
Same manufacturer · Stryker Endoscopy
See all →- ModerateStryker PneumoSure Insufflators Recalled for Flow Rate Drop
FDA (Devices) · 2016-05-04
- SevereStryker Recalls SERFAS 90 Degree Energy Probes Due to Fragmentation
FDA (Devices) · 2015-07-01
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Same hazard · documentation error
See all →- LowMindray cPM Series Patient Monitor Manuals Recalled for Alarm Clarification
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- LowMindray Recalls Passport Series Patient Monitor Operator's Manuals
FDA (Devices) · 2026-10-14
- ModerateConMed Suture Anchor Recalled for Incorrect Patient Implant Card
FDA (Devices) · 2026-09-02
- ModerateHenry Schein Recalls Structur 2 SC Temporary Resin Cartridges for Incorrect Packaging
FDA (Devices) · 2026-09-02
Similar recalls
- ModerateStryker Arthroscope and Hardware Set Tray Recall for Sterilization Labeling
FDA (Devices) · 2012-07-18
- ModerateStryker AccuPlace Round Level Needle Guide Recalled for Sterility Breach Risk
FDA (Devices) · 2016-03-09
- ModerateStryker Irrigator Reusable Tips Recalled for Incorrect Sterilization Instructions
FDA (Devices) · 2013-05-15
- ModerateStryker 12-Instrument Sterilization Tray Recalled for Validation Testing Gaps
FDA (Devices) · 2014-05-14
- ModerateStryker AccuPlace Drace Stereotaxic Needle Guide Recalled for Packaging Sterility Risk
FDA (Devices) · 2016-03-09
- ModerateStryker AccuPlace Needle Guide Recalled for Sterility Breach Risk
FDA (Devices) · 2016-03-09