Stryker Hoffmann LRF Transport Strut Locking Mechanism Recall
Stryker is recalling Hoffmann LRF Transport Strut devices due to oversized thread pitch that prevents the distal ring fixation from locking, affecting 488 units nationwide.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device recall for a functional defect (oversized thread pitch) that prevents proper locking of an orthopedic fixation device. No injuries or hospitalizations reported; the defect creates a risk of harm through device malfunction but no adverse events have been documented.
Plain-English summary
Stryker is recalling specific lots of the Hoffmann LRF Bone Transport Strut (Model 4933-0-400), an orthopedic external fixation device intended to keep bone fragments in proper position during healing. The recalled devices have an oversized thread pitch at the distal end of the threaded rod, which prevents the distal ring fixation from locking between 0–50mm of the threaded rod.
This defect affects 488 units that were distributed nationwide. The affected devices are identified by specific serial numbers listed in the recall notice. Because the fixation ring cannot lock properly, the device cannot perform its intended function of maintaining correct positioning of bone fragments during the healing process.
Patients and healthcare providers with affected serial numbers should contact Stryker for guidance regarding affected devices.
The recalled product
- Product
- stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achieved
- Manufacturer
- Stryker GmbH
- Affected units
- 488
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial numbers: D32158
- D32160
- D32161
- D32162
- D32163
- D32164
- D34343
- D34344
- D40690
- D41629
- D41630
- D41631
- D44713
- G16331
- G16332
- G16333
- G26278
- G29759
- G29760
- G32730
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Stryker GmbH
See all →- HighStryker Hoffmann II Carbon Connecting Rods Recalled Over Potential Cracking After Sterilization
FDA (Devices) · 2026-08-05
- HighStryker Hoffmann II Connecting Rod Micro May Crack After Cleaning and Sterilization
FDA (Devices) · 2026-08-05
- HighStryker Hoffmann II Carbon Connecting Rods Recalled Over Potential Cracking After Sterilization
FDA (Devices) · 2026-08-05
- HighStryker Hoffmann II Carbon Connecting Rods Recalled Over Potential Cracking After Sterilization
FDA (Devices) · 2026-08-05
- HighStryker Hoffmann II Carbon Connecting Rods Recalled Over Potential Cracking
FDA (Devices) · 2026-08-05
Same hazard · device malfunction
See all →- SevereMedtronic MiniMed Flex Insulin Pumps Recalled for Failed Verification Tests
FDA (Devices) · 2026-10-14
- ModerateWilson-Cook Recalls Hemospray Endoscopic Hemostat Due to Adherence Issues
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Et Control Software Recalled for Anesthetic Delivery Interruption
FDA (Devices) · 2026-09-23
Similar recalls
- SevereStryker Recalls Hoffman LRF Hexapod Struts Due to Pin Displacement Risk
FDA (Devices) · 2018-11-07
- SevereStryker Recalls Hoffmann LRF Hexapod Struts Due to Pin Fit Defect
FDA (Devices) · 2018-11-07
- SevereStryker Recalls Hoffman LRF Telescopic Struts Due to Breakage
FDA (Devices) · 2012-11-14
- HighStryker Recalls Hoffmann LRF Safety Clips Due to Loosening Risk
FDA (Devices) · 2017-05-31
- HighStryker Hoffman LRF Wire Tensioner Adjustment Ring May Become Loose
FDA (Devices) · 2024-07-03
- ModerateStryker Recalls 18 Hinge Coupling Units Due to Packaging Error
FDA (Devices) · 2018-05-02