Medical Device RayStation Dosage Calculation Inconsistency in Radiation Therapy Planning
RayStation radiation therapy treatment planning software contains an inconsistency in how it handles density uncertainty during dose calculations for proton and light ion treatments, potentially affecting treatment planning accuracy.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device recall without reported injuries or hospitalizations. The inconsistency in dose calculation poses a potential risk of harm to treatment planning accuracy, but no patient injury has been reported in the source. This meets the High severity criterion for risk-of-harm medical devices without reported injury.
Plain-English summary
RayStation is a radiation therapy treatment planning system manufactured by RaySearch Laboratories. Versions 10.1.0.613 (RayStation 10B) and 10.1.1.54 (10B Service Pack 1) contain an inconsistency in how the software handles density uncertainty when calculating treatment doses for proton and light ion therapies.
The problem affects three software functions: Robust optimization, Robust evaluation, and Compute perturbed dose. When using a HU-to-mass density CT (Hounsfield Unit) calibration curve, the software does not consistently account for density uncertainty, which could lead to incorrect dose calculations during treatment planning.
This recall affects radiation therapy treatment planning facilities worldwide, including locations in the United States, Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, Switzerland, Taiwan, Thailand, and the United Kingdom.
Healthcare facilities using affected versions of RayStation should contact RaySearch Laboratories for software updates and verification procedures for treatment plans created with the affected software.
The recalled product
- Product
- Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1
- Manufacturer
- RAYSEARCH LABORATORIES AB
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 11 recalled by RAYSEARCH LABORATORIES AB under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 11 products in this recall →Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28