The Recall Desk

FDA (Devices) recall action 96156

RAYSEARCH LABORATORIES AB recalled 11 products on 2025-02-26

The FDA (Devices) publishes a recall like this one as 11 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 11 pages.

Products
11
Announced
2025-02-26
Critical
0
Units
46

Why these products were recalled

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–11 of 11

  • HighFDA (Devices)··2025-02-26

    Medical device recall: RayStation radiation therapy software dose calculation inconsistency

    RayStation radiation therapy treatment planning software contains an inconsistency in density uncertainty calculations that may affect dose accuracy for proton and light ion treatments. The issue affects systems using HU-to-mass density CT calibration curves from RayStation versions 4.5 through 2024B.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 5.0.2.35 and 5.0.3.17 Software Version: RayStation 5 Service Pack 2 and 3
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    RayStation Radiation Therapy Planning Software Recall for Density Calculation Inconsistency

    RayStation radiation therapy planning software contains an inconsistency in how it calculates density uncertainty for proton and light ion treatments. This affects versions 4.5 through 2024B and could impact treatment plan accuracy.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Version: RayStation 12A, 12A Service Pack 1 and 2 and 2023B
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    RayStation radiation therapy software: incorrect density uncertainty calculations in dose planning

    RayStation radiation therapy planning software versions 4.5 through 2024B contains an inconsistency in density uncertainty calculations affecting proton and light ion dose computations. No illnesses reported.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Version: RayStation 11B and 11B Service Pack 1, 2, 3 and Service Pack Toshiba 1
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    Radiation Therapy Planning Software Correction for Density Uncertainty in Dose Calculations

    RayStation radiation therapy planning software contains an inconsistency in density uncertainty calculations within dose planning functions affecting proton and light ion treatments in versions 4.5 through 2024B.

    Product
    Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Service Pack 1
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    Medical Device RayStation Dosage Calculation Inconsistency in Radiation Therapy Planning

    RayStation radiation therapy treatment planning software contains an inconsistency in how it handles density uncertainty during dose calculations for proton and light ion treatments, potentially affecting treatment planning accuracy.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    RayStation radiation therapy software recall for density calculation inconsistency

    RayStation treatment planning software versions 4.5-2024B contain a calculation error in density uncertainty functions affecting proton and light ion treatment plans. The error could lead to incorrect dose calculations.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.0.0.24 Software Version: RayStation 6
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    RayStation radiation therapy planning software density uncertainty calculation inconsistency

    RayStation radiation therapy planning software contains an inconsistency in density uncertainty calculations affecting robust optimization and dose computation functions for proton and light ion treatment plans.

    Product
    Brand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52 and 10.0.2.10 Software Version: RayStation 9B, 9B Service Pack 1, 10A, 10A Service Pack 1 and 2
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    RayStation radiation therapy planning software density uncertainty computational defect

    RayStation treatment planning software versions 4.5 through 2024B may incorrectly handle density uncertainty in radiation dose calculations for proton and light ion therapy. An inconsistency in three key functions could affect treatment planning accuracy.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.1.1.2, 6.2.0.7, 6.3.0.6 Software Version: RayStation 6 Service Pack 1, 2 and 3
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    Radiation Therapy Planning System Density Calculation Inconsistency in RayStation

    RayStation radiation therapy planning software versions 4.5 through 2024B contain an inconsistency in how density uncertainty is handled in dose calculation functions for proton and light ion therapy plans.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 1.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15 Software Revisions: RayStation 11A and 11A Service Pack 1, 2 and 3
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    RayStation radiation therapy planning software density uncertainty calculation inconsistency

    RayStation radiation therapy planning software versions 4.5 through 2024B contain an inconsistency in density uncertainty handling when HU-to-mass density CT calibration curves are used. No illnesses or injuries have been reported.

    Product
    Brand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113 and 9.0.1.142 Software Version: RayStation 8B, 8B Service Pack 1 and 2, 9A and 9A Service Pack 1
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-02-26

    RayStation Radiation Therapy Software Recalled for Dose Calculation Inconsistency

    RayStation radiation therapy planning software is being recalled due to inconsistent handling of density uncertainty in dose calculations for proton and light ion treatment plans. The software defect may affect the accuracy of calculated radiation doses.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4 and 4.7.6.7 Software Version: RayStation 4.5.1, 4.7.2, 4.7.3, 4.7.4, 4.7.5 and 4.7.6
    Category
    Distribution
    0 states