The Recall Desk
HighFDA (Devices)·Z-1170-2015·Announced 2015-03-04

Navilyst NAMIC Convenience Kits Recalled for Sterility Concerns

Navilyst Medical is recalling 411 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (infection risk due to potential sterility breach) where injury has not yet been reported, fitting the High severity criteria.

Plain-English summary

Navilyst Medical, Inc. is recalling 411 units of NAMIC Convenience Kits (UPN H749601314811, REF/Catalog No. 60131481). These sterile, prescription-only devices are intended for use in Fluid Management and/or Invasive Pressure Monitoring Systems. The recall covers lots 4245021 through 4822657, as listed in the source data.

The recall was initiated because the labeled expiration dating of the affected product is not fully supported by shelf life verification testing. This deficiency may result in a sterility breach. Use of a Fluid Management Convenience Kit where sterility has been breached may result in an increased risk of infection.

The affected products were distributed worldwide, including US Nationwide, Canada, and Guam. Consumers and healthcare providers should stop using the affected lots and contact Navilyst Medical for further instructions.

The recalled product

Product
NAMIC Convenience Kit, UPN H749601314811, REF/Catalog No. 60131481, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
Affected units
411

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Lots 4245021
  • 4260368
  • 4268542
  • 4279333
  • 4309640
  • 4344841
  • 4380268
  • 4387481
  • 4390493
  • 4413788
  • 4436076
  • 4436659
  • 4447611
  • 4474955
  • 4491914
  • 4503486
  • 4505497
  • 4534914
  • 4549037
  • 4573406

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 60 recalled by Navilyst Medical, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 60 products in this recall →