The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1156-2015·Announced 2015-03-04

Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

Navilyst Medical is recalling 927 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in sterility breach and increased infection risk.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (sterile medical device) where the hazard is an increased risk of infection due to potential sterility breach, but no specific injuries or illnesses have been reported in the source text.

Plain-English summary

Navilyst Medical, Inc. is recalling 927 units of the NAMIC Convenience Kit, 3 VALVE MANIFOLD KIT (Catalog No. 60041161). These sterile, single-use devices are intended for use in Fluid Management and/or Invasive Pressure Monitoring Systems. The recall covers specific lots distributed nationwide in the US, as well as in Canada and Guam.

The recall was initiated because the labeled expiration dating of the affected product is not fully supported by shelf life verification testing. This deficiency may result in a sterility breach. The use of a Fluid Management Convenience Kit where sterility has been breached may result in an increased risk of infection.

Healthcare providers and facilities should stop using the affected lots immediately and contact Navilyst Medical for instructions on return or disposal. Consumers should not use these products if they have them in their possession.

The recalled product

Product
NAMIC Convenience Kit, 3 VALVE MANIFOLD KIT, UPN H749600411611, REF/Catalog No. 60041161, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure…
Affected units
927

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Lots 4581678
  • 4588087
  • 4595774
  • 4597572
  • 4620890
  • 4623072
  • 4638352
  • 4638441
  • 4638716
  • 4649828
  • 4666871
  • 4667422
  • 4672695
  • 4673930
  • 4675943
  • 4682904
  • 4683074
  • 4690302
  • 4691995
  • 4696832

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 60 recalled by Navilyst Medical, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 60 products in this recall →

Similar recalls