The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1123-2013·Announced 2013-04-24

Advanced Sterilization Products Recalls EVOTECH Endoscope Cleaner Systems

Advanced Sterilization Products is recalling 273 EVOTECH Endoscope Cleaner & Reprocessor Systems because they did not receive sufficient High Potential (HIPOT) DC voltage testing.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a lack of electrical safety testing (HIPOT), which poses a risk of harm (electrical shock/fire) even though no injuries have been reported yet, fitting the 'High' criteria for risk-of-harm products where injury has not yet been reported.

Plain-English summary

Advanced Sterilization Products (ASP) has determined that a limited number of EVOTECH Endoscope Cleaner & Reprocessor Systems (P/N 50004) manufactured between 2011 and 2012 did not receive sufficient High Potential (HIPOT) DC voltage testing. The recall involves 273 units that were distributed nationwide in the United States, including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.

The EVOTECH system is designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes. The specific defect identified is the lack of sufficient High Potential (HIPOT) DC voltage testing, which is a standard safety test for electrical equipment. The source text does not report any illnesses, injuries, or deaths associated with this recall.

Healthcare facilities and users should check their inventory for the specific serial numbers listed in the official recall notice. If a unit is found, it should be removed from service and the manufacturer contacted for instructions on repair or replacement.

The recalled product

Product
EVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes.
Affected units
273

Distribution

Distributed nationwide across the United States.

The notice also names 44 states specifically: